Fluke 975 Meter Users Manual Supplement

This document is a supplement to the Fluke 975 Users Manual. It contains information necessary to ensure the accuracy of the manual.


Fluke 975 Meter Users Manual Supplement - cover page
Brand
Fluke
Category
Other
Document type
Other
Language
English
Pages
8
File format
PDF
File size
500 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
BC64A27691A3A05CA86926BE0FD3B546

About this manual

This is Issue 7 of the Fluke 975 Users Manual Supplement, part number 2507213, issued 10/11 and printed August 2006. The supplement contains information needed to keep the 975 Users Manual accurate.

It replaces the specifications on pages 27 through 30 and lists general, calculated, environmental, and physical specifications, plus agency approvals. It also adds calibration gas concentrations and regulator requirements, replacement part numbers, and updated calibration setup and replaceable parts figures.

What's inside

  • Change #1, 46008
  • General Specifications
  • Calculated Parameters
  • Environmental Specifications
  • Physical Specifications
  • Agency Approvals, Certifications and Standards Compliance
  • Change #2
  • Change #3
  • Calibration Set up
  • Replaceable Part and Accessories

Specifications

Relative humidity range10 to 90 % R.H. non-condensing
Air velocity range50 to 3000 fpm (0.25 to 15 m/sec)
CO2 range0 to 5000 ppm
CO range0 to 500 ppm
Meter weight0.544 kg (1.2 lb)
Meter size28.70 cm x 11.43 cm x 5.08 cm (11.3 in x 4.5 in x 2 in)
Operating and storage temperature-20 to 50 °C (-4 to 122 °F)
AltitudeUp to 2000 m (6562 ft)
Rechargeable Lithium battery life11 hours at room temperature
Backup AA battery life7 hours

Frequently asked questions

What is the operating temperature range of the Fluke 975?

The meter operates and stores from -20 to 50 °C (-4 to 122 °F).

What CO and CO2 concentrations are used for calibration?

Calibration gas is specified as CO at 200 ppm nominal (150 to 250 ppm) and CO2 at 5000 ppm nominal (4500 to 5000 ppm).

What flow rate does the calibration regulator require?

The regulator requirement is 0.5 liters per minute.

How long does the rechargeable Lithium battery last?

It provides 11 hours of operation at room temperature.