The EcoSense EcoSense® ODO200 ® ODO200M Dissolved Oxygen & Temperature Instrument User Manual describes the features and specifications of the product.
The YSI EcoSense ODO200/ODO200M is a handheld instrument that measures temperature and dissolved oxygen in % and mg/L. It uses optical, lifetime luminescence technology and a digital signal between instrument and probe, and it is powered by one 9V battery. The manual covers package contents, probe preparation, battery installation, keypad and display, measurement modes, saving and viewing data, and calibration.
The ODO200 stores 50 data sets, while the ODO200M stores 250 data sets and adds a real-time clock, a waterproof USB port for downloading data to a PC, and a user-selectable recalibration interval. The manual also covers sensor maintenance and storage, troubleshooting, error messages, specifications, accessories, and contact information.
| Temperature Range | 0 to 50 °C |
|---|---|
| Temperature Resolution and Accuracy | 0.1 °C resolution; ±0.3 °C accuracy |
| Dissolved Oxygen Range (% air saturation) | 0.0 to 200% air saturation |
| Dissolved Oxygen Range (mg/L) | 0.00 to 20.0 mg/L |
| Pressure Compensation Input | 500 to 1125 mBar (375 to 844 mmHg) |
| Salinity Compensation Input | 0.0 to 40.0 ppt |
| Power Source | One 9V battery (included with meter) |
| Battery Life (ODO200) | 100 hours in manual sampling mode |
| Battery Life (ODO200M) | 52 hours in manual sampling mode |
| Memory | 50 data sets (ODO200); 250 data sets (ODO200M) |
| Ingress Protection | IP-67 with cable connector cap installed |
| Weight (with battery) | 281 grams (.62 lbs) |
| Recal Timer | 0 to 60 days (ODO200M only) |
The ODO200 stores 50 data sets, and the ODO200M stores 250 data sets.
The instrument uses one 9V battery.
The average sensor cap replacement interval is 12 to 18 months, although caps may last longer.
The ODO200M features a waterproof USB port with cover for downloading stored measurement data to a PC.