Siemens Sysmex CA-600 Series Coagulation Analyzers Technical Specifications (CA-620 / CA-660)

The Sysmex®CA-600 Systems incorporate the proven technology and reliability of the Sysmex®CA-500 Series Systems, making them easy and safe to integrate into your laboratory.


Siemens Sysmex CA-600 Series Coagulation Analyzers Technical Specifications (CA-620 / CA-660) - cover page
Brand
Siemens
Category
Analyzers
Document type
Product Leaflet
Language
Spanish
Pages
2
File format
PDF
File size
595 KB
Published
15 November, 2023
Updated
15 November, 2023
MD5 checksum
C7D895F413D0122B49B3EE8BAC0D8DC1

About this manual

This two-page document is the Spanish technical specification sheet (ficha de especificaciones técnicas) for the Siemens Sysmex CA-600 series coagulation analyzers. The series builds on the proven technology of the Sysmex CA-500 series and comes in two configurations: the CA-620 for coagulation detection only, and the CA-660 for coagulation, chromogenic and immunological detection methods. Key features include a 2D barcode reader, a tilted color LCD touch screen, and automated positive sample identification via barcode.

The sheet lists the assay menu (PT, APTT, fibrinogen, thrombin time, reptilase time, D-dimer, protein C, heparin, antithrombin, lupus anticoagulant and von Willebrand factor assays, some available only on the CA-660), followed by detailed specifications covering reaction cuvettes, hardware and software, measurement principle, sample and reagent handling, throughput, quality control, data storage, environmental conditions and ordering information. The systems are available for sale in the U.S. and comply with the IVD Directive 98/79/EC. This manual is written in Spanish.

What's inside

  • Menú de análisis
  • Especificaciones de los Sistemas de la Serie Sysmex CA-600
  • Componentes de hardware y software
  • Control de calidad
  • Medición
  • Manejo de muestras
  • Rendimiento
  • Manejo de reactivos
  • Operatividad
  • Información de pedido

Specifications

System optionsCA-620 (coagulation); CA-660 (coagulation, chromogenic, immunological)
Measurement principlePhoto-optical, continuous, sequential
Methods/channelsCoagulation/4, chromogenic and immunological
Reading lampsCoagulation: 4 LEDs at 660 nm; chromogenic: 1 LED at 405 nm; immunological: 1 LED at 575 nm
Reaction cuvettesIndividual; 60 cuvettes on board with automatic continuous access
WorkstationBuilt-in PC with built-in touch screen
PrinterBuilt-in thermal graphic printer
InterfaceASTM protocol, CA-1000; bidirectional RS-232C serial host connection
Power supply100 to 240 VAC; 50-60 Hz; 300 VA; dedicated line required
Operating conditions15-35°C (optimal 23°C); 30-85% humidity, non-condensing
Dimensions / weight566 x 490 x 490 mm (W x D x H); approx. 43 kg
Sample handlingPrimary tubes and sample cups; automated sampling with standard predilution; max 10 sample tubes
ThroughputPT 60 tests/hour; PT/APTT/Fibrinogen 48 simultaneous tests/hour; D-dimer 16 tests/hour
Reagents on board13 positions (including buffer and priming solution); 1 thermostated dispensing pipette
Data storage600 samples / 3,000 patient results; QC: 14 parameters, 180 pts x 6 files per parameter
Catalog numbersCA-620: 10712040; CA-660: 10712039

Frequently asked questions

What is the difference between the Sysmex CA-620 and CA-660?

The CA-620 performs coagulation detection only, while the CA-660 also supports chromogenic and immunological detection methods.

Which assays can the CA-600 series run?

The menu includes PT, APTT, fibrinogen, thrombin time, reptilase time, factor deficiency, lupus anticoagulant, protein C, heparin, antithrombin, D-dimer and von Willebrand factor assays; some, such as D-dimer and vWF, are available only on the CA-660.

How fast is the CA-600 series?

Throughput is approximately 60 PT tests per hour, 48 simultaneous PT/APTT/fibrinogen tests per hour, and 32 simultaneous PT/APTT/fibrinogen/D-dimer tests per hour.

What power supply does the CA-600 require?

It runs on 100 to 240 VAC at 50-60 Hz, consumes 300 VA, and requires a dedicated power line.