The Sysmex®CA-600 Systems incorporate the proven technology and reliability of the Sysmex®CA-500 Series Systems, making them easy and safe to integrate into your laboratory.
This two-page document is the Spanish technical specification sheet (ficha de especificaciones técnicas) for the Siemens Sysmex CA-600 series coagulation analyzers. The series builds on the proven technology of the Sysmex CA-500 series and comes in two configurations: the CA-620 for coagulation detection only, and the CA-660 for coagulation, chromogenic and immunological detection methods. Key features include a 2D barcode reader, a tilted color LCD touch screen, and automated positive sample identification via barcode.
The sheet lists the assay menu (PT, APTT, fibrinogen, thrombin time, reptilase time, D-dimer, protein C, heparin, antithrombin, lupus anticoagulant and von Willebrand factor assays, some available only on the CA-660), followed by detailed specifications covering reaction cuvettes, hardware and software, measurement principle, sample and reagent handling, throughput, quality control, data storage, environmental conditions and ordering information. The systems are available for sale in the U.S. and comply with the IVD Directive 98/79/EC. This manual is written in Spanish.
| System options | CA-620 (coagulation); CA-660 (coagulation, chromogenic, immunological) |
|---|---|
| Measurement principle | Photo-optical, continuous, sequential |
| Methods/channels | Coagulation/4, chromogenic and immunological |
| Reading lamps | Coagulation: 4 LEDs at 660 nm; chromogenic: 1 LED at 405 nm; immunological: 1 LED at 575 nm |
| Reaction cuvettes | Individual; 60 cuvettes on board with automatic continuous access |
| Workstation | Built-in PC with built-in touch screen |
| Printer | Built-in thermal graphic printer |
| Interface | ASTM protocol, CA-1000; bidirectional RS-232C serial host connection |
| Power supply | 100 to 240 VAC; 50-60 Hz; 300 VA; dedicated line required |
| Operating conditions | 15-35°C (optimal 23°C); 30-85% humidity, non-condensing |
| Dimensions / weight | 566 x 490 x 490 mm (W x D x H); approx. 43 kg |
| Sample handling | Primary tubes and sample cups; automated sampling with standard predilution; max 10 sample tubes |
| Throughput | PT 60 tests/hour; PT/APTT/Fibrinogen 48 simultaneous tests/hour; D-dimer 16 tests/hour |
| Reagents on board | 13 positions (including buffer and priming solution); 1 thermostated dispensing pipette |
| Data storage | 600 samples / 3,000 patient results; QC: 14 parameters, 180 pts x 6 files per parameter |
| Catalog numbers | CA-620: 10712040; CA-660: 10712039 |
The CA-620 performs coagulation detection only, while the CA-660 also supports chromogenic and immunological detection methods.
The menu includes PT, APTT, fibrinogen, thrombin time, reptilase time, factor deficiency, lupus anticoagulant, protein C, heparin, antithrombin, D-dimer and von Willebrand factor assays; some, such as D-dimer and vWF, are available only on the CA-660.
Throughput is approximately 60 PT tests per hour, 48 simultaneous PT/APTT/fibrinogen tests per hour, and 32 simultaneous PT/APTT/fibrinogen/D-dimer tests per hour.
It runs on 100 to 240 VAC at 50-60 Hz, consumes 300 VA, and requires a dedicated power line.