Teledyne LeCroy HDO6000/HDO6000A High Definition Oscilloscopes Operator's Manual

This manual describes the features and specifications of Teledyne LeCroy oscilloscopes.


Teledyne LeCroy HDO6000/HDO6000A High Definition Oscilloscopes Operator's Manual - cover page
Brand
Teledyne Lecroy
Category
Other Electronics
Document type
User Manual
Language
English
Pages
210
File format
PDF
File size
16.5 MB
Published
11 November, 2023
Updated
11 November, 2023
Model
HDO6000
MD5 checksum
E32E0F26F25F69714D3232A407A61AE8

About this manual

This operator's manual covers the operation and maintenance of all instruments in the Teledyne LeCroy MAUI HDO6000 series of High Definition Oscilloscopes, with the HDO6000A shown in the illustrations. It explains the front, side and back of the instrument, front panel controls, signal interfaces, probes, powering on and off, software activation and connecting to other devices and systems.

The manual then walks through the MAUI touch screen interface and working with traces, followed by chapters on acquisition, timebase, triggering, display, cursors, measurements, math functions, memory, analysis tools such as WaveScan, PASS/FAIL testing and the spectrum analyzer, saving and recalling data, utilities, and maintenance procedures including restart, restore default setup, touch screen calibration and firmware updates.

What's inside

  • Oscilloscope Overview and Set Up
  • Using MAUI
  • Acquisition
  • Trigger
  • Cursors
  • Measure
  • Math
  • Analysis Tools
  • Saving Data (File Functions)
  • Utilities and Maintenance

Specifications

Operating temperature5 °C to 40 °C
Operating humidityMaximum 90% RH up to 31 °C, decreasing linearly to 50% RH at 40 °C
Operating altitudeUp to 10,000 ft (3,048 m) at or below 30 °C
AC power100-240 Vrms (±10%) at 50/60 Hz (±5%); 100-120 Vrms (±10%) at 400 Hz (±5%)
Maximum power consumption320 W (320 VA)
Nominal power consumption200 W (200 VA)
Standby power consumption4 W
Maximum input voltage (50 Ω coupling)≤ 5 Vrms
Maximum input voltage (1 MΩ coupling)≤ 400 Vpk max (Peak AC ≤ 10 kHz + DC), derating above 10 kHz

Frequently asked questions

Does the HDO6000 require manual line voltage selection?

No. The instrument automatically adapts to the line voltage; manual voltage selection is not required.

How should I clean the oscilloscope?

Clean only the exterior using a soft cloth moistened with water or an isopropyl alcohol solution, and unplug the power cord before cleaning. Do not use harsh chemicals or attempt to clean internal parts.

How much clearance does the instrument need for cooling?

In a benchtop configuration, leave a minimum of 15 cm (6 inches) around the sides between the instrument and the nearest object, and do not block the cooling vents.

Can the measuring terminals be connected directly to supply mains?

No. The measuring terminals have no rated measurement category per IEC/EN 61010-1:2010 and are not intended to be connected directly to supply mains.

Download PDF (16.5 MB)