The pressure balance is used to measure pressure and has the characteristics of high precision, high stability and high reliability.
The WIKA CPB5800 is a pressure balance used for calibrating electronic or mechanical pressure measuring instruments. It uses a piston-cylinder measuring system and can be fitted with single-range or dual-range piston-cylinder systems depending on the measuring range.
Operation includes setting up the device, installing the piston-cylinder system, venting, loading masses, approaching the pressure value and releasing pressure. The manual also covers safety instructions, troubleshooting, cleaning, consumable parts, recalibration, specifications, tables of masses and accessories. The accuracy is 0.015% as standard, optionally 0.006% of reading, and the recommended recalibration interval is 2 to 5 years depending on conditions of usage.
| Accuracy | 0.015% as standard (optional 0.006%) of reading |
|---|---|
| Recommended recalibration interval | 2 to 5 years |
| Tank volume | 250 ml |
| Standard gravity | 9.80665 m/s² |
| Warranty period | 24 months |
| Measuring ranges, single-range | 120 bar and 300 bar (or 1,600 psi and 4,000 psi) |
| Measuring ranges, dual-range | 700 bar, 1,200 bar and 1,400 bar (or 10,000 psi, 16,000 psi and 20,000 psi) |
The CPB5800 has an accuracy of 0.015% as standard, with an optional accuracy of 0.006% of reading.
The recommended recalibration interval is 2 to 5 years, depending on conditions of usage.
The warranty period is 24 months according to the general terms of supply of ZVEI.
The masses are manufactured to standard gravity of 9.80665 m/s² and can be adjusted for a customer's location or gravity value.