The TP-AD-04 digestion furnace adopts a quartz tube infrared heating digestion method, which is characterized by a large digestion heating area, small temperature difference, fast temperature rise, good sample digestion consistency and high thermal efficiency.
The TP-AD-04 digestion furnace from Zhejiang Top Instrument is used to shorten sample digestion time when determining protein content. It heats by quartz tube infrared radiation, which the manual describes as giving a large heated area, small temperature differences, fast warm-up, consistent digestion and high thermal efficiency, and it is stated to meet the CE standard. A smart microcontroller controls the heating process, so after the heating mode, the three temperature stages and the heating times are set, digestion runs automatically without an attendant. Harmful gases such as SO2 leaving the digestion tubes are drawn off through the collection tube on the furnace by a suction pump and discharged with water into the drain.
The manual explains how the suction pump is connected to the tap and the drain collection tube, the functions of the panel keys, how the heating stages and temperatures are set, and the step-by-step operating procedure. It also lists precautions on gas leakage, the water supply and cooling, and gives the accessory list supplied with the instrument. This manual is written in Chinese.
| Heating method | Quartz tube infrared heating |
|---|---|
| Number of temperature stages | 3 (T1, T2, T3) |
| Recommended stage settings | T1 280 °C for 20 min; T2 450 °C for 60–90 min; T3 350 °C for 0 min |
| Water flow rate for suction pump | 3–4 L per minute |
| Typical digestion time | 1.5–2 hours from power-on to complete digestion |
| Cooling time before switching off | 30 minutes after digestion completes |
| Standard | CE |
A typical run takes 1.5 to 2 hours from switch-on to complete digestion.
Yes; a stage that is not needed has its heating time set to 0.
The instrument must cool for 30 minutes after digestion before the power is turned off.