CCS51 Free Chlorine Sensor BA01905C/28/ZH/03.20

Compact design, easy to install Stainless steel material, strong corrosion resistance Stable and reliable output signal, high accuracy Long service life, low maintenance cost


CCS51 Free Chlorine Sensor BA01905C/28/ZH/03.20 - cover page
Brand
Endress+Hauser
Category
Industrial Supply
Document type
User Manual
Language
Chinese
Pages
45
File format
PDF
File size
1.8 MB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
D99678640F3F1157A4F0422E02CC197E

About this manual

The Endress+Hauser CCS51 free chlorine sensor is designed for measuring free chlorine concentration in drinking water, process water, and bathing water. It operates on a membrane method principle, utilizing a gold working electrode and a silver/antimony counter electrode to generate a current proportional to the free chlorine concentration. The sensor includes a built-in temperature sensor and features a robust PVDF membrane with a pressure relief valve. This manual is written in Chinese.

What's inside

  • Document Information
  • Basic Safety Guidelines
  • Product Description
  • Receiving Inspection
  • Installation
  • Electrical Connection
  • Commissioning
  • Diagnostics and Troubleshooting
  • Maintenance
  • Repair and Disposal

Frequently asked questions

What is the CCS51 sensor used for?

It continuously measures free chlorine concentration in water and is used with measuring and control equipment to optimize disinfection control.

What minimum flow velocity is required at the sensor membrane?

The flow velocity at the membrane must not be lower than 15 cm/s (0.5 ft/s).

What volumetric flow corresponds to the minimum flow velocity when the sensor is installed in a Flowfit CCA151 assembly?

It is 5 l/h (1.3 gal/h).

What is the highest medium pH allowed for calibrating the complete measuring system?

The medium pH must not be higher than pH 8.