Hanna Instruments HI98302 TDS Meter Instruction Manual

The HI98302 is an economical TDS meter that features a waterproof housing, large backlit display, and automatic temperature compensation. It is ideal for water quality monitoring, aquariums, and hydroponics applications.


Hanna Instruments HI98302 TDS Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
Spanish
Pages
2
File format
PDF
File size
235 KB
Published
06 November, 2023
Updated
06 November, 2023
MD5 checksum
4B329F0FD91DBE791F48E1D47A4489F2

About this manual

Hanna Instruments HI98302 is a handheld TDS meter with a 0.00 to 10.00 ppt measurement range and automatic temperature compensation from 0.0 to 50.0 °C. The manual covers measurement, single-point automatic calibration at 6.44 ppt, error messages, battery replacement, and setup options.

It also lists specifications such as 0.01 ppt resolution, CR2032 battery life of about 250 hours, and operating conditions, plus warranty terms of one year for the meter and three months for probes. This manual is written in Spanish.

What's inside

  • Garantía
  • Soluciones
  • Accesorios
  • Cuidado y mantenimiento
  • Mensajes de error
  • Procedimiento de calibración
  • Medición y calibración
  • Cambio de la pila
  • Especificaciones
  • Guía de uso

Specifications

TDS range0.00 to 10.00 ppt
Temperature range0.0 to 50.0 °C (32.0 to 122.0 °F)
TDS resolution0.01 ppt
Temperature resolution0.1 °C (0.1 °F)
Accuracy±2% F.S.; ±0.5 °C (±1.0 °F)
TDS conversion factor0.5
CalibrationSingle point, automatic at 6.44 ppt
Temperature compensationAutomatic from 0.0 to 50.0 °C (32.0 to 122.0 °F)
BatteryCR2032 lithium 3 V, 1 unit
Battery lifeApproximately 250 hours continuous use
Operating conditions0 to 50 °C (32 to 122 °F); max 100% RH
Weight68 g without battery
Recommended immersion depth3 cm
Instrument warranty1 year
Probe warranty3 months
Default temperature scale°C
Automatic shutoff options8 minutes, 60 minutes, or disabled

Frequently asked questions

How do I calibrate the HI98302?

Enter calibration mode, pour 5 cm of standard solution, center and immerse the probe 3 cm; the meter shows REC until stable, then Stor.

What does the '---- Err' message mean?

It appears during user calibration when the reading is outside the accepted range or the solution is not recognized.

How do I replace the battery?

Turn the battery cap counterclockwise, remove it, and insert a CR2032 with the positive side up.

What are the measurement ranges?

TDS from 0.00 to 10.00 ppt and temperature from 0.0 to 50.0 °C.