JOFRA ASM-801/802/803 A/B Advanced Signal Multi-scanner Reference Manual

This is a user manual, which introduces how to use AMETEK multi-scanner


JOFRA ASM-801/802/803 A/B Advanced Signal Multi-scanner Reference Manual - cover page
Brand
JOFRA
Category
Calibration Components
Document type
User Manual
Language
English
Pages
62
File format
PDF
File size
752 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
E503A17CD8E5C2C32D5879C927D2A664

About this manual

The JOFRA ASM-801/802/803 A/B Advanced Signal Multi-scanner is a calibration instrument designed for industrial temperature calibration. It supports RTDs, thermocouples, voltage sources, and 4-20 mA transmitters, with multiple scan rates and remote operation via RS-232.

The manual provides safety instructions, installation steps, sensor connection details, operating procedures (both manual and with JOFRACAL software), remote command set, maintenance guidelines, error codes, adjustment procedures using AMETRIM software, technical specifications, and a list of accessories.

What's inside

  • Introduction
  • Safety instructions
  • Receiving the multi-scanner
  • Multi-scanner models (A/B versions)
  • Setting up the multi-scanner
  • Operating the multi-scanner
  • Remote operation (RS-232 commands)
  • Maintenance
  • Errors
  • Adjusting the multi-scanner (AMETRIM software)

Specifications

Current input signal range0 – 24 mA
Current input internal power supply24 V, max. 28 mA
Current input display resolution0.0001 mA / 0.001 °C / 0.001 °F
Current input accuracy±(0.010% of reading + 0.010% of full scale)
Current input temperature coefficient7 ppm/°C
Current input impedance<10 Ω
Voltage input signal range0 – 12 V
Voltage input display resolution0.0001 V / 0.001 °C / 0.001 °F
Voltage input accuracy±(0.005% of reading + 0.010% of full scale)
Voltage input temperature coefficient5 ppm/°C
Voltage input impedance>1 MΩ
Power supply input90 – 254 VAC, 45 – 65 Hz
Power supply output30 V ±2% regulated DC, max. 30 W
Communication interfaceRS232
Transmission rate115200 Baud
Scan rates (B-model)Scan rate 1: 1 ch/sec (V, mA, mV), 0.5 ch/sec (RTD); Scan rate 2: 2 ch/sec (V, mA, mV), 1 ch/sec (RTD); Scan rate 3: 3 ch/sec (V, mA, mV), 1.5 ch/sec (RTD); Scan rate 4: 5 ch/sec (V, mA, mV), 2.5 ch/sec (RTD)
Weight2.3 kg / 5.07 lb
Dimensions (L×W×H)250 × 249 × 69 mm / 9.8 × 9.8 × 2.7 in
Operating temperature0 – 40 °C / 32 – 104 °F
Storage temperature-20 – 50 °C / -4 – 122 °F
Humidity range0–90% RH
Protection classIP30
Supported RTD sensor typesCUSTOM ITS-90, CUSTOM CvD, P10(90)385, P50(90)385, P100(90)385, P200(90)385, P500(90)385, P1K(90)385, P100(90)392, H120(90)672, M50(90)428, M100(90)428, YSI(90)400
Supported thermocouple typesB, E, J, K, L, N, R, S, T, U, BP, XK

Frequently asked questions

What is the maximum input voltage allowed for the multi-scanner?

Do not connect the multi-scanner to input voltages above 30 Volts.

What communication interface and baud rate does the multi-scanner use for remote control?

The standard communication interface is RS232 at 115200 Baud.

What is the operating temperature range of the multi-scanner?

The operating temperature range is 0 – 40 °C (32 – 104 °F).