JOFRA ETC-400 R Infrared Temperature Calibrator Specification Sheet

Specification sheet for the JOFRA ETC-400 R infrared temperature calibrator: 28–400°C range, fast heating, stability indicator, auto-stepping, and IR-LAB…


JOFRA ETC-400 R Infrared Temperature Calibrator Specification Sheet - cover page
Brand
JOFRA
Category
Calibration Kit
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
346 KB
Published
09 November, 2026
Updated
09 November, 2026
MD5 checksum
118B7629D4B005778D27388E44CB835D

About this manual

The JOFRA ETC-400 R is a dry-block temperature calibrator designed for calibrating infrared thermometers in the field. It covers 28 to 400°C / 82 to 752°F, heats up as quickly as 100°C per minute, and stabilizes in about 3 minutes. Its 36 mm target has an emissivity of 0.96 and a coating designed for long-term performance under high temperature.

Operation uses one-key-one-function keys, a backlit multi-information display, a stability indicator with a 3-minute countdown, and auto-stepping with up to 9 programmable temperature steps. Standard delivery includes an RS232 interface, AMECAL-LIGHT calibration software, and JOFRA IR-LAB software for correcting emission-factor settings.

What's inside

  • Product Description
  • Temperature Ranges
  • Fast Calibration
  • Extreme Flexibility
  • Stability Indicator
  • Auto-Stepping
  • Maximum Temperature
  • Instrument Setups
  • Re-calibration
  • JOFRA IR-LAB Software

Specifications

Temperature Range28 to 400°C / 82 to 752°F at 23°C ambient
ResolutionSelectable 1° or 0.1°
Heating Time (28 to 200°C)2 minutes
Heating Time (200 to 400°C)3 minutes
Cooling Time (400 to 200°C)9 minutes
Stability±0.3°C / ±0.54°F
Accuracy±0.5°C / ±0.9°F
Time to StabilityApproximately 3 minutes
Mains Voltage115V (90-127) or 230V (180-254)
Frequency45 - 65 Hz
Power Consumption (max.)350 W
Target Emissivity0.96

Frequently asked questions

What is the temperature range of the JOFRA ETC-400 R?

The ETC-400 R covers 28 to 400°C / 82 to 752°F at an ambient temperature of 23°C / 73°F.

What software is included with the ETC-400 R?

Standard delivery includes the AMECAL-LIGHT calibration software and JOFRA IR-LAB software for calibrating IR thermometers with a fixed emission factor setting.

How long does the ETC-400 R take to heat up?

It heats from 28 to 200°C in about 2 minutes and from 200 to 400°C in about 3 minutes, reaching stability in approximately 3 minutes.

What is the emissivity of the ETC-400 R calibration surface?

The calibration surface of the ETC-400 R has an emission factor of 0.96.

Download PDF (346 KB)

More JOFRA Manuals


View all JOFRA manuals →