This document is the user manual and reference manual for the Temperature Calibrator
This user- and reference manual covers the JOFRA ETC-125 A, ETC-400 A and ETC-400 R temperature calibrators. It is divided into 10 chapters that describe how to set up, operate, service and maintain the calibrator, and it includes technical specifications and a list of accessories.
Safety symbols in the manual identify warnings, cautions and notes. The document also states that thermal specifications are given at an ambient temperature of 23°C/73.4°F ± 3°C/5.4°F.
| Maximum temperature (ETC-125 A) | 125°C / 257°F |
|---|---|
| Minimum temperature (ETC-125 A) | -18°C / -0.4°F at ambient 0°C / 32°F; -10°C / 14°F at ambient 23°C / 73.4°F; 6°C / 42.8°F at ambient 40°C / 104°F |
| Display resolution | 1° or 0.1°C / 1.8° or 0.18°F |
| Stability | ±0.05°C / ±0.09°F |
| Accuracy | ±0.5°C / ±0.9°F |
| Power supply (ETC-125 A) | Multivoltage 115VAC and 230VAC, 47-63Hz; 115 V (90-132) and 230 V (180-264) |
| Power consumption (ETC-125 A) | 75 VA |
| Power supply (ETC-400 A / R) | 115VAC, 45-65Hz: 90-127; 230VAC, 45-65Hz: 180-254 |
| Power consumption (ETC-400 A / R) | 350 W |
| Weight (ETC-125 A) | 1.75 kg / 3.86 lb |
| Dimensions | 172 x 72 x 182 mm / 6.8 x 2.8 x 7.2 inch |
| Operating temperature | 0 to 40°C / 32 to 104°F |
| Protection class | IP10 |
All thermal specifications are given with an ambient temperature of 23°C/73.4°F ± 3°C/5.4°F.
The manual cites EN 61326-1: 2013 for EMC requirements.
The protection class is IP10.
For the ETC-125 A, heating from 23 to 100°C / 73 to 212°F takes 11 minutes.