JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual

This document is the user manual and reference manual for the Temperature Calibrator


JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual - cover page
Brand
JOFRA
Category
Test Set
Document type
User Manual
Language
English
Pages
84
File format
PDF
File size
965 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
8BD070FC4476B9C1BE9C25C4B9B263F1

About this manual

This user- and reference manual covers the JOFRA ETC-125 A, ETC-400 A and ETC-400 R temperature calibrators. It is divided into 10 chapters that describe how to set up, operate, service and maintain the calibrator, and it includes technical specifications and a list of accessories.

Safety symbols in the manual identify warnings, cautions and notes. The document also states that thermal specifications are given at an ambient temperature of 23°C/73.4°F ± 3°C/5.4°F.

What's inside

  • About this manual
  • Safety symbols
  • User manual
  • Reference manual
  • Technical specifications
  • List of accessories

Specifications

Maximum temperature (ETC-125 A)125°C / 257°F
Minimum temperature (ETC-125 A)-18°C / -0.4°F at ambient 0°C / 32°F; -10°C / 14°F at ambient 23°C / 73.4°F; 6°C / 42.8°F at ambient 40°C / 104°F
Display resolution1° or 0.1°C / 1.8° or 0.18°F
Stability±0.05°C / ±0.09°F
Accuracy±0.5°C / ±0.9°F
Power supply (ETC-125 A)Multivoltage 115VAC and 230VAC, 47-63Hz; 115 V (90-132) and 230 V (180-264)
Power consumption (ETC-125 A)75 VA
Power supply (ETC-400 A / R)115VAC, 45-65Hz: 90-127; 230VAC, 45-65Hz: 180-254
Power consumption (ETC-400 A / R)350 W
Weight (ETC-125 A)1.75 kg / 3.86 lb
Dimensions172 x 72 x 182 mm / 6.8 x 2.8 x 7.2 inch
Operating temperature0 to 40°C / 32 to 104°F
Protection classIP10

Frequently asked questions

What ambient temperature is used for the thermal specifications?

All thermal specifications are given with an ambient temperature of 23°C/73.4°F ± 3°C/5.4°F.

Which EMC standard is cited for the calibrators?

The manual cites EN 61326-1: 2013 for EMC requirements.

What is the protection class of the ETC-125 A and ETC-400 A/R?

The protection class is IP10.

What is the heating time of the ETC-125 A from 23 to 100°C?

For the ETC-125 A, heating from 23 to 100°C / 73 to 212°F takes 11 minutes.

Download PDF (965 KB)

More JOFRA Manuals


View all JOFRA manuals →