This reference manual covers the setup, operation, maintenance, and technical specifications for the Jofra CTC-320/650 B and CTC-1200 A temperature calibrators.
This document serves as a reference guide for the Jofra CTC-320/650 B and CTC-1200 A temperature calibrators, produced by AMETEK Denmark A/S. It is designed for users familiar with AMETEK calibrators as well as new users, providing instructions on setup, operation, service, and maintenance. The manual includes safety instructions, detailed operating procedures for features like the SWITCH TEST and AUTO STEP, and guidance on storing and transporting the instrument.
| Max. Temperature (CTC-320 B) | 320°C/608°F |
|---|---|
| Max. Temperature (CTC-650 B) | 650°C (1202°F) |
| Max. Temperature (CTC-1200 A) | 1205°C (2201°F) |
| Min. Temperature | 50°C/122°F |
| Display resolution | 0.1°C/ 0.1°F |
| Stability | ±0.1°C/±0.18°F |
| Accuracy | ±0.5°C/±0.9°F |
The maximum temperature for the CTC-320 B model is 320°C/608°F.
The minimum operating temperature is 50°C/122°F.
New insertion tubes must be heated up to maximum temperature for a period of minimum 30 minutes.