testo 476 frequency tachometer operation manual. This manual provides detailed information on the product's features, characteristics, and usage
The testo 476 is a handheld stroboscopic tachometer designed for industrial, laboratory, R&D, and academic environments. It makes fast-moving objects appear slower or stationary, allowing non-contact measurement of rotational or linear speed. The device operates with one hand and includes a battery charger (100–240 V AC), external trigger, transport case, and this manual.
The manual details basic safety warnings, intended use, scope of delivery, instrument overview, power supply and battery management, operating steps, error messages (e.g., E1 for front ring not secured), external trigger specifications (DTL/TTL compatible, 0 to +5 V square wave), stroboscopic tube replacement procedure, and application information such as slowing motion, determining apparent direction, harmonics, and finding true RPM. This manual is written in Chinese.
| Measurement unit | Flashes per minute (FPM) = RPM |
|---|---|
| External trigger compatibility | DTL/TTL compatible, detects square wave 0 to +5 V, minimum 800 nsec pulse width |
| Battery charger input | 100–240 V AC |
| Low battery operation time | Approximately 4 minutes after low battery indicator appears |
| Charging time | Approximately 2–3 hours |
Insert the cylindrical plug into the charger end and the charger socket on the testo 476, then plug the charger into a wall outlet. The red LED indicates charging; it flashes when fully charged. Charging takes about 2–3 hours.
Error E1 indicates the front ring is not properly secured. Turn off the instrument, remove the front ring, reset the reflector, and re-secure the front ring.
The external trigger is DTL/TTL compatible, detects a square wave signal from 0 to +5 V with a minimum pulse width of 800 ns. The 1/8" jack provides common ground, +7.2 V DC output for sensors, and the center connection is the trigger input.