testo 476 Stroboscopic Tachometer Operation Manual

testo 476 frequency tachometer operation manual. This manual provides detailed information on the product's features, characteristics, and usage


testo 476 Stroboscopic Tachometer Operation Manual - cover page
Brand
testo
Category
Test Set
Document type
User Manual
Language
Chinese
Pages
13
File format
PDF
File size
419 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
D17F6C886A7D73FAC7864E7E1370E630

About this manual

The testo 476 is a handheld stroboscopic tachometer designed for industrial, laboratory, R&D, and academic environments. It makes fast-moving objects appear slower or stationary, allowing non-contact measurement of rotational or linear speed. The device operates with one hand and includes a battery charger (100–240 V AC), external trigger, transport case, and this manual.

The manual details basic safety warnings, intended use, scope of delivery, instrument overview, power supply and battery management, operating steps, error messages (e.g., E1 for front ring not secured), external trigger specifications (DTL/TTL compatible, 0 to +5 V square wave), stroboscopic tube replacement procedure, and application information such as slowing motion, determining apparent direction, harmonics, and finding true RPM. This manual is written in Chinese.

What's inside

  • Basic Safety Instructions
  • Intended Use
  • Scope of Delivery
  • Overview
  • Power Supply
  • Operation
  • Error Messages
  • External Trigger
  • Replacing Stroboscopic Tube
  • Application Information

Specifications

Measurement unitFlashes per minute (FPM) = RPM
External trigger compatibilityDTL/TTL compatible, detects square wave 0 to +5 V, minimum 800 nsec pulse width
Battery charger input100–240 V AC
Low battery operation timeApproximately 4 minutes after low battery indicator appears
Charging timeApproximately 2–3 hours

Frequently asked questions

How do I charge the testo 476 battery?

Insert the cylindrical plug into the charger end and the charger socket on the testo 476, then plug the charger into a wall outlet. The red LED indicates charging; it flashes when fully charged. Charging takes about 2–3 hours.

What does error E1 mean and how do I fix it?

Error E1 indicates the front ring is not properly secured. Turn off the instrument, remove the front ring, reset the reflector, and re-secure the front ring.

What are the external trigger specifications?

The external trigger is DTL/TTL compatible, detects a square wave signal from 0 to +5 V with a minimum pulse width of 800 ns. The 1/8" jack provides common ground, +7.2 V DC output for sensors, and the center connection is the trigger input.