This document is a manual supplement, which provides additional information about the 5900E user manual
This document is a manual supplement for the Fluke 5900E Users Guide, issued in 2018 by Fluke Corporation. It contains specific changes to ensure the accuracy of the original manual, covering updates to the mercury triple point cell specifications, care instructions, and realization techniques.
| ITS-90 Assigned Temperature | 234.3156 K (–38.8344 °C) |
|---|---|
| Expanded Uncertainty | 0.2 mK |
| Purity of Mercury | 99.999999+% (8N+) |
| Quantity of Mercury | 2,600 g |
| Outer Diameter of the Cell | 38 mm (1.5 in) |
| Length of the Cell | 400 mm (15.75 in) |
| Inner Diameter of the Reentrant Well | 8 mm (0.31 in) |
| Total Immersion Depth in Mercury | 210 mm (8.27 in) |
| Total Length of the Reentrant Well | 381 mm (15 in) |
| Case Material | 304L Stainless Steel |
Fluke Calibration recommends that most users realize the triple point of mercury using the melting technique.
The realization of the melting curve of the mercury triple point is accomplished in a bath with a minimum depth of 457 mm (18 inch).
The immersion depth of the midpoint of a SPRT sensor in a mercury triple point cell is approximately 183 mm.