Teledyne 3190-MT Micro Oxygen Analyzer Operation Manual

This manual introduces the basic operation and maintenance knowledge of the Teledyne 3190-MT micro oxygen analyzer, including the components, working principles, installation, operation and maintenance of the instrument.


Teledyne 3190-MT Micro Oxygen Analyzer Operation Manual - cover page
Brand
Teledyne
Category
Analyzers
Document type
User Manual
Language
Chinese
Pages
58
File format
PDF
File size
571 KB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
92D1B1785A3E0EAEEDFB3105C1A5D592

About this manual

Teledyne 3190-MT is a microprocessor-based trace oxygen analyzer for measuring trace oxygen in inert gases or mixed gases. It uses a micro fuel cell sensor and provides two user-defined ranges from 0-10 ppm to 0-9999 ppm, two adjustable concentration alarms, a sensor failure alarm, RS-232 output, and three analog outputs.

The manual covers unpacking, installation, wiring, gas connections, operation, calibration, fuse replacement, and sensor replacement. The analyzer has an 8-bit CMOS microprocessor with onboard RAM and 16 KB ROM, AC (100-240 V, 50/60 Hz) or DC (10-36 V) power, and a compact enclosure with a flowmeter. This manual is written in Chinese.

What's inside

  • Introduction
  • Operating Principle
  • Micro Fuel Cell Sensor
  • Electrical Principle
  • Installation
  • Operation
  • Maintenance
  • Appendix

Specifications

Oxygen measurement range0-10 ppm to 0-25% oxygen
User-defined rangesTwo ranges from 0-10 ppm to 0-9999 ppm
AC power100-240 V AC, 50/60 Hz
DC power10-36 V DC
Analog outputs0-10 V DC concentration, 0-10 V DC range ID, 4-20 mA DC concentration (negative ground)
CommunicationRS-232 serial port
Microprocessor8-bit CMOS microprocessor
MemoryOnboard RAM and 16 KB ROM
Sensor dimensionsDiameter 1 1/4 in; height 1 1/4 in
Cathode areaApproximately 4 cm2
Sensor failure alarm thresholdAbove 4.447 ppm/nA
Concentration alarmsTwo adjustable alarm points
Sensor warranty12 months
Air calibration20.9% oxygen
Oxidizing gasesNot affected by oxidizing gases

Frequently asked questions

What gases can the 3190-MT measure?

It measures trace oxygen in inert gases or various mixed gases.

What AC power does the analyzer require?

100-240 V AC at 50/60 Hz.

What DC voltage is required?

10 to 36 V DC.

When does the sensor failure alarm trigger?

When the ratio exceeds 4.447 ppm/nA.

Download PDF (571 KB)

More Teledyne Manuals


View all Teledyne manuals →