testo 410-2 wind speed meter operation manual, mainly introduces the use methods and precautions of the product.
This Chinese operation manual covers the Testo 410-2, a handheld instrument that measures air velocity, humidity and temperature and is commonly used to measure airflow at ventilation duct outlets. It also states that the instrument can be used to analyze climatic conditions.
The manual walks through the parts of the instrument, inserting the two AAA batteries, the configuration mode for units and auto-off, selecting the display parameter and view (Hold, Max, Min, Hold Avg), time-averaged airflow calculation, resetting Max/Min values, cleaning the housing and the common Hi/Lo display messages. A technical data section lists measuring ranges, resolution, accuracy, protection class, battery life, dimensions and weight. This manual is written in Chinese.
| Measurement parameters | m/s, km/h, fpm, mph, kts, Beaufort, °C, wind chill temperature, %RH, wet bulb, dew point |
|---|---|
| Measuring range | 0.4 to 20 m/s, 80 to 4000 fpm, -10 to +50 °C, 0 to 100 %rH |
| Resolution | 0.1 m/s, 1 fpm, 0.1 °C, 0.1 %rH |
| Accuracy | ±(0.2 m/s + 2% of reading), ±(40 fpm + 2% of reading), ±0.5 °C, ±2.5 %rH (5 to 95 %rH) |
| Measuring rate | 0.5 s; humidity 1 s |
| Protection class | IP 10 |
| Ambient temperature | -10 to +50 °C |
| Storage/transport temperature | -40 to +70 °C |
| Power supply | 2 x 1.5 V AAA batteries |
| Battery life | 80 hours (backlight off) |
| Dimensions | 133 x 46 x 25 mm (including protective cap) |
| Weight | 110 g (including batteries and protective cap) |
Air velocity from 0.4 to 20 m/s (80 to 4000 fpm), temperature from -10 to +50 °C and humidity from 0 to 100 %rH.
It uses 2 x 1.5 V AAA batteries, with a battery life of 80 hours when the backlight is off.
The reading is outside the measuring range, or less than 10 minutes of battery power remain, in which case the batteries should be replaced.
Press the button until Hold and Avg appear, then hold it until ---- appears and release to start averaging; press the button to end the measurement and show the mean value.