victor3900 is a special standard resistor for lightning protection. It is composed of 9 1/2w high-precision military-grade wire-wound resistors. It can be directly switched to test the resistance when calibrating the resistance. It does not need to change the test line. It has stable performance, sturdy structure, and convenient calibration testing.
VICTOR 3900 is a lightning protection standard resistor used to calibrate ground resistance testers and equipotential testers. It consists of nine 1/2 W high-precision military-grade wirewound resistors and provides resistance values from 0.001 Ω to 100 kΩ.
The manual describes its technical specifications, instrument structure, operating method, and packing list. It states that calibration must use the four-wire method and that the resistor must not be subjected to more than 1/2 W. This manual is written in Chinese.
| Function | Calibration of ground resistance testers and equipotential testers |
|---|---|
| Construction | 9 high-precision military-grade 1/2 W wirewound resistors |
| Resistance ranges and accuracy | 0.001 Ω (2%), 0.01 Ω (0.5%), 0.1 Ω (0.2%), 1.0 Ω (0.2%), 10 Ω (0.05%), 100 Ω (0.05%), 1 kΩ (0.05%), 10 kΩ (0.05%), 100 kΩ (0.05%) |
| Connection method | Four-wire method |
| Temperature coefficient | ±25 ppm |
| Dimensions | 173 mm x 125 mm x 70 mm |
| Mass | 360 g |
| Test leads | Red 0.5 m and black 0.5 m, 2 each |
| Operating temperature and humidity | -10 to 40 °C; below 70% RH |
| Storage temperature and humidity | -20 to 60 °C; below 80% RH |
| Insulation resistance | Above 10 MΩ (circuit to housing, 500 V) |
| Withstand voltage | AC 3700 V/rms (circuit to housing) |
| Safety compliance | IEC 61010-1, CAT III 600 V, pollution degree, JJG166-1993 |
Calibration must use the four-wire method; otherwise measurement accuracy cannot be guaranteed.
Do not apply more than 1/2 W, or the standard resistor may be damaged.
The packing list includes one instrument, four test leads, one instrument bag, and one user manual with warranty card and certificate.