BCHP GCK3302/GCK3308 Air Source User Manual

This is a manual of BCHP GCK3302/GCK3308 air source, which introduces the name of each part of the instrument, installation and use, working principle and characteristics, main technical parameters and other information.


BCHP GCK3302/GCK3308 Air Source User Manual - cover page
Brand
BCHP
Category
Air Cleaners Purifiers
Document type
User Manual
Language
Chinese
Pages
12
File format
PDF
File size
324 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
6067D0EC1FBA4815AFD96B29429E621B

About this manual

The BCHP GCK3302/GCK3308 automatic air source is a pressure-stable, purified air source for gas chromatographs. It can serve evaporative light scattering detectors or multiple gas chromatographs and replaces high-pressure air cylinders.

It uses a two-stage pressure control system with pressure stability within 0.002 MPa, an automatic water discharge circuit, and an electronic low-voltage start system. The manual covers installation, operation, working principle, technical parameters, accessories, guarantee, and troubleshooting. This manual is written in Chinese.

What's inside

  • Company introduction
  • Product overview
  • Instrument parts and names
  • Installation and use
  • Working principle and features
  • Main technical parameters
  • Instrument completeness
  • Manufacturer's guarantee
  • Troubleshooting

Specifications

Output flow, GCK33025000 ml/min at 0.4 MPa
Output flow, GCK33088000 ml/min at 0.4 MPa
Output pressure0-0.4 MPa
Pressure stability<0.002 MPa
Working noise<35 dB(A)
Working environment0-45 C, relative humidity <85%
Power consumption300 W
Working voltage220 V +/-10%
Dimensions480 x 260 x 350 mm
Net weight25 kg

Frequently asked questions

What are the output flows of the GCK3302 and GCK3308?

The GCK3302 has an output flow of 5000 ml/min and the GCK3308 has 8000 ml/min, both at 0.4 MPa.

How often should the filter adsorbent be replaced?

The adsorbent in the filter must be replaced every three months.

What does it mean if pressure drops more than 0.02 MPa during self-check?

It indicates a leak; check the instrument for leaks and see the troubleshooting section.

Download PDF (324 KB)

More BCHP Manuals