BCHP GCK3302/GCK3308 Air Source User Manual

User Manual 12 pages · 324 KB · ZH

The BCHP GCK3302/GCK3308 automatic air source is a pressure-stable, purified air source for gas chromatographs. It can serve evaporative light scattering detectors or multiple gas chromatographs and replaces high-pressure air cylinders.

It uses a two-stage pressure control system with pressure stability within 0.002 MPa, an automatic water discharge circuit, and an electronic low-voltage start system. The manual covers installation, operation, working principle, technical parameters, accessories, guarantee, and troubleshooting. This manual is written in Chinese.

Read online

BCHP GCK3302 specifications

Output flow, GCK33025000 ml/min at 0.4 MPa
Output flow, GCK33088000 ml/min at 0.4 MPa
Output pressure0-0.4 MPa
Pressure stability<0.002 MPa
Working noise<35 dB(A)
Working environment0-45 C, relative humidity <85%
Power consumption300 W
Working voltage220 V +/-10%
Dimensions480 x 260 x 350 mm
Net weight25 kg

Source: BCHP GCK3302/GCK3308 Air Source User Manual

All BCHP GCK3302 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What are the output flows of the GCK3302 and GCK3308?

The GCK3302 has an output flow of 5000 ml/min and the GCK3308 has 8000 ml/min, both at 0.4 MPa.

From: BCHP GCK3302/GCK3308 Air Source User Manual

How often should the filter adsorbent be replaced?

The adsorbent in the filter must be replaced every three months.

From: BCHP GCK3302/GCK3308 Air Source User Manual

How often should the activated carbon filter be replaced?

The activated carbon in the filter must be replaced every three months, or approximately after 500 hours of operation.

From: BCHP GCK3302/GCK3308 Fully Automatic Air Source User Manual

What indicates a leak during the self-check procedure?

If the compressor restarts within 30 minutes after the initial startup, it indicates an air leak in the instrument.

From: BCHP GCK3302/GCK3308 Fully Automatic Air Source User Manual