BCHP fully automatic air source GCK3302/GCK3308 is a pressure-stable and pure air source specifically designed for gas chromatographs. The product features high pressure stability and a pressure stability accuracy better than 0.002MPa.
The BCHP GCK3302 and GCK3308 are fully automatic air sources designed specifically for gas chromatographs. They serve as a stable and pure compressed air supply, replacing high-pressure air cylinders. The units feature a two-stage pressure stabilization control system with a stability precision better than 0.002MPa. This manual is written in Chinese.
| Output Flow (GCK3302) | 0-5000ml/min at 0.4MPa |
|---|---|
| Output Flow (GCK3308) | 0-8000ml/min at 0.4MPa |
| Output Pressure | 0-0.4MPa |
| Pressure Stability | <0.002MPa |
| Operating Noise | <35dB(A) |
| Power Consumption | 250W |
| Operating Voltage | 220V±10%V |
The activated carbon in the filter must be replaced every three months, or approximately after 500 hours of operation.
If the compressor restarts within 30 minutes after the initial startup, it indicates an air leak in the instrument.
The output flow for the GCK3302 model ranges from 0 to 5000ml/min at a state of 0.4MPa.