Testo 510 Differential Pressure Meter User Manual

Testo 510 is a device for measuring differential pressure. It is often used for measuring small differential pressure (such as detecting the permeability of filters), airflow pressure, small suction and pitot tube wind speed.


Testo 510 Differential Pressure Meter User Manual - cover page
Brand
testo
Category
Other device
Document type
User Manual
Language
Chinese
Pages
8
File format
PDF
File size
229 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
DF21A1D1A8CD9F4664E5EBC8CFA943D2

About this manual

Testo 510 is a differential pressure meter used for measuring small differential pressure, airflow pressure, small suction, and pitot tube air velocity. It supports multiple pressure and velocity units and has a measurement range of 0 to 100 hPa.

The manual covers safety instructions, specifications, product description, initial setup, operation, maintenance, and troubleshooting. It describes battery installation, basic settings, zeroing, display views, reset functions, averaging, and instrument care. This manual is written in Chinese.

What's inside

  • Safety instructions
  • Specifications
  • Product description
  • Overview
  • Initial steps
  • Instrument operation
  • Maintenance
  • Tips and help
  • FAQ

Specifications

SensorDifferential pressure sensor
Measurement parametersPa, hPa, mbar, mmH2O, mmHg, inHG, inH2O, psi, m/s, fpm
Measurement range0 to 100 hPa; 0 to 40.15 inH2O
Resolution0.01 hPa; 0.01 inH2O
Accuracy±0.03 hPa (0–0.30 hPa), ±0.05 hPa (0.31–1.00 hPa), ±(1.5% reading + 0.1 hPa) (remaining range); ±0.01 inH2O (0–0.12 inH2O), ±0.02 inH2O (0.13–0.40 inH2O), ±(1.5% reading + 0.04 inH2O) (remaining range)
Protection ratingIP 40
Ambient temperature0 to +50 °C (+32 to +122 °F)
Storage/transport temperature-40 to +70 °C (-40 to +158 °F)
Power supply2 x 1.5 V AAA batteries
Battery life50 hours (backlight off)
Dimensions119 x 46 x 25 mm
Weight90 g (including battery and protective cap)
EC standard89/336/EEC
Measurement rate0.5 s

Frequently asked questions

What is the Testo 510 used to measure?

It is used for small differential pressure, airflow pressure, small suction, and pitot tube air velocity measurement.

What is the measurement range of the Testo 510?

The measurement range is 0 to 100 hPa, or 0 to 40.15 inH2O.

How long do the batteries last?

Battery life is 50 hours with the backlight off.

What does Hi or Lo on the display mean?

It can mean the reading is outside the measurement range, or the remaining battery life is less than 10 minutes; replace the battery if needed.

Download PDF (229 KB)

More testo Manuals


View all testo manuals →