The product is used to measure the sulfur content of combustible materials such as coal, coke and oil. The entire testing process is automatically completed by the microcomputer and can also be done without the microcomputer. The experiment is easy to calibrate and intuitive. The measurement results can be stored for a long time and can be queried and printed.
This manual describes a microcomputer sulfur detector used to measure sulfur content in combustible materials such as coal, coke, and petroleum by coulometric titration. It explains the working principle, system components, installation, software operation, experimental procedure, fault diagnosis, reagents, and packing list.
The instrument combines a high-temperature furnace, electrolytic cell, magnetic stirrer, air purification unit, automatic sample feeder, and microcontroller system. The manual also provides technical parameters, correction methods, electrolyte preparation, sample preparation, and an installation diagram. This manual is written in Chinese.
| Sulfur measurement range | 0.01-40% |
|---|---|
| Carrier gas flow | 1000 ml/min |
| Sulfur determination time | 3-9 min, automatically controlled |
| Heating time | 20-45 min |
| Power supply | Single-phase AC ~220 V, 50 Hz, 2 kW |
| Accuracy | Conforms to GB/T 214 |
| Selectable furnace temperature | 600-1150 C (coal 1050 C; oil 800 C selectable) |
| Electrolytic cell volume | About 400 ml |
| Stirrer speed | About 500 rpm |
| Air flow meter range | 0-1000 ml/min |
The new overall coefficient equals the old overall coefficient plus (standard value minus actual value) divided by the actual value.
Insert the program disc into the drive, open the drive, and click the SetupNoName.exe icon; installation completes automatically.