PSE575/576/577 general fluid pressure sensor user manual, mainly introduces the product safety precautions, type representation, product parts name and use precautions, installation, wiring methods, fault list, specifications and other contents.
This user manual covers the PSE575/576/577 general fluid pressure sensor. It presents safety precautions, usage conditions, warranty and limitation notes, and model designation information.
The manual then explains part names and usage precautions, installation and setup, wiring, a troubleshooting table, specifications and outline dimensions. This manual is written in Chinese.
| Pressure sensor type | Ceramic diaphragm pressure sensor |
|---|---|
| Tightening torque for R1/4 thread | 8 to 12 N·m |
| Maximum wiring length | 10 m or less |
| Cable tensile strength | 50 N or less |
| Recommended cable bend radius | Larger of 6 times the sheath outer diameter or 33 times the insulation outer diameter |
The manual specifies 8 to 12 N·m for the R1/4 thread.
Wiring should be as short as possible and no longer than 10 m.
No. The manual states that the product cannot be used as a legal measuring instrument.
Use the larger of 6 times the sheath outer diameter or 33 times the insulation outer diameter.