Testo 420 Flow Hood Instruction Manual

This manual describes the operation of the testo 420 flow hood.


Testo 420 Flow Hood Instruction Manual - cover page
Brand
testo
Category
Other
Document type
User Manual
Language
English
Pages
26
File format
PDF
File size
3.4 MB
Published
29 October, 2023
Updated
29 October, 2023
MD5 checksum
8A5E0D41F59B242947ECC764BEE83475

About this manual

The Testo 420 is used for volume flow measurements as its main application, plus Pitot tube measurements and pressure measurements for air conditioning and ventilation systems. Its interchangeable flow hoods allow use on air inlets and outlets of various sizes, and an Android/iOS App can display readings on a tablet or smartphone and start, stop and save a measurement.

The instruction manual covers safety and environmental protection, specifications, product description, first steps and operation. It explains Bluetooth settings, measurement settings such as damping, hood calibration and zeroing interval, volume flow measurement setup, measuring, saving, transferring data to a PC, cleaning, troubleshooting, accessories and authorizations.

What's inside

  • Safety and the environment
  • Specifications
  • Product description
  • First steps
  • Using the product
  • Volume flow measurement setup
  • Measuring
  • Saving
  • Transferring measurement data to the PC
  • Maintaining the product

Specifications

Bluetooth range>20 m (free field)
Measurement cycle1/sec
Volume flow measuring range40 to 4000 m³/h / 25 to 2300 cfm / 11 to 1100 l/s
Operating temperature-5 to +50 °C/+23 - +122°F
Weight of entire measurement setupapprox. 2900 g
Battery lifetimeapprox. 40 h
Standard hood size610x610 mm
Measuring instrument dimensions150x85x35 mm

Frequently asked questions

What measurements can the Testo 420 perform?

It is used for volume flow measurements as the main application, Pitot tube measurements and pressure measurements for air conditioning and ventilation systems.

How long does the battery last?

The battery lifetime is approximately 40 hours with a zeroing interval of 10 seconds, display illumination off and Bluetooth off.

What size is the standard hood?

The standard hood is 610x610 mm.

Can mild household cleaning agents be used?

Yes, mild household cleaning agents or soap suds may be used.