This document is the user manual for the blood glucose and ketone monitoring system. It mainly introduces how to use the blood glucose and ketone monitoring system, including the system composition, settings, and usage methods.
This user manual explains how to operate the Abbott Precision Xtra Blood Glucose and Ketone Monitoring System, including setup, testing procedures, interpreting results, and troubleshooting. It covers both blood glucose and blood ketone monitoring, with detailed instructions for using test strips and managing the meter’s memory and settings. The manual also includes safety information, care instructions, and specifications for the device.
The system is designed for home or professional use in managing diabetes and measures β-hydroxybutyrate (ketone) in fingertip blood. It requires specific Abbott test strips and includes a lancing device, logbook, and carrying case. Users must follow instructions carefully to ensure accurate results and avoid infection risks when testing multiple patients.
| Size | 7.47 cm (L) x 5.33 cm (W) x 4.32 cm (H) |
|---|---|
| Weight | 40-46 grams |
| Power Source | One CR 2032 Lithium (coin cell) battery |
| Battery Life | Approximately 1,000 tests |
| Memory | Up to 450 events |
| System Altitude | Up to 2195 meters (7200 feet) above sea level |
The Display Check shows on the display window briefly each time you turn the meter on. Do not use the meter if the Display Check does not exactly match the picture in the manual, as it may show incorrect results.
Use only Precision Xtra® Blood Glucose Test Strips and Precision Xtra® Blood β-Ketone Test Strips. Other test strips may produce inaccurate results.
Gently push the battery cover in and up, lift it out, remove the old battery, insert a new CR 2032 Lithium battery with the plus sign facing up, and snap the cover back into place.
The meter can store up to 450 events, including control solution results, personal blood glucose and blood ketone results, and other meter information.