This is an operating manual, which mainly introduces the pump's characteristics, indications for use, precautions, panel displays, control dial settings, pump sets, instructions for use, battery operation, checking pump accuracy, troubleshooting, cleaning and other contents
The Abbott PATROL Enteral Pump is a microprocessor-controlled rotary peristaltic pump designed to provide accurate, controlled enteral feeding. It features a large display, simple controls, and a bolus-prevention feature for safe administration of liquid enteral feeding products, infant formulas, or reconstituted powder solutions.
| Flow rate range | 1 to 300 mL/hr in 1-mL/hr increments |
|---|---|
| Flow rate accuracy | +10% or +0.5 mL/hr, whichever is greater |
| Battery operation time | 8 hours at 125 mL/hr |
| Maximum dose setting | 9999 mL |
| Volume accumulation limit | 9999 mL |
The pump allows a flow rate selection from 1 to 300 mL/hr in 1-mL/hr increments.
The pump offers 8 hours of battery operation at a flow rate of 125 mL/hr when fully charged.
No, this device is strictly for enteral use only and must not be used for intravenous (IV) administration.