PerkinElmer 263A Power Booster User's Manual

This manual describes the hardware functionality and usage instructions for PerkinElmer instruments.


PerkinElmer 263A Power Booster User's Manual - cover page
Brand
PerkinElmer
Category
medical technology
Document type
User Manual
Language
English
Pages
14
File format
PDF
File size
2.9 MB
Published
27 October, 2023
Updated
27 October, 2023
MD5 checksum
7A5C0145D951D27CF3CB0B49AAFFC839

About this manual

This manual provides instructions for using the PerkinElmer 263A Power Booster, an add-on system for electrochemical applications. It details how to connect the booster to a potentiostat and KEPCO power supply, configure settings, and perform initial checks. The manual includes safety warnings and specifications for different models (263A/10A, 263A/20A, 263A/8A).

What's inside

  • General
  • System Specifications
  • Interconnections
  • Connecting the Potentiostat to the Power Booster
  • 263A Cell Cable Connections
  • KEPCO Front-Panel Setup
  • External Cell Connection and Initial Check

Specifications

Rated Output Current20 A (nominal) for the 263A/20A system; 10 A for the 263A/10A system; and 8 A (nominal) for the 263A/8A system.
Rated Output Voltage20 V at 20 A (nominal) for the 263A/20A and 263A/10A; 50 V at 8 A (nominal) for the 263A/8A.
Bandwidth5 KHz at 3 db with 1 V applied at counter with 1-/G36 cell (1-W load)
Voltage Rise Time<70 µs (10–90%, measured at Counter, ± 10 V/10-Hz square wave into a 1-W load, high-speed mode)
Potential Noise<500 µV rms at E Monitor Out
Current Measurement Rise Time<75µs (10–90%, measured at I Monitor Out, ±10 A/10-Hz square wave)
Current Measurement Accuracy±1% of reading, ± 20 mA
Applied Potential Accuracy±0.5% of reading, ±2 mV (nominal; determined by the KEPCO)

Frequently asked questions

What are the different models of the 263A Power Booster?

The 263A/10A provides ±10 amps (nominal) at ±20 volts (nominal) compliance. The 263A/20A provides ±20 amps (nominal) at ±20 volts (nominal) compliance. The 263A/8A provides ±8 amps (nominal) at ±50 volts (nominal) compliance.

How do I connect the 263A to the KEPCO power supply?

Connect the S-COM brown wire to S COM on the KEPCO rear panel and to screw-clamp connector 3 on the Power Booster Interface. Connect the S-OUT red wire to S OUT on the KEPCO rear panel and to screw-clamp connector 4 on the Power Booster Interface. Use the cardedge cable to connect the E/IN and I/OUT connectors.

What is the current measurement accuracy?

The current measurement accuracy is ±1% of reading, ± 20 mA.

What is the applied potential accuracy?

The applied potential accuracy is ±0.5% of reading, ±2 mV (nominal; determined by the KEPCO).