This document is a user guide for PerkinElmer ViewLux™ CCD Imager, it introduces how to set up the optical elements, instrument parameters and how to run LanthaScreen® Terbium Assay
This document provides a setup guide for the PerkinElmer ViewLux™ CCD Imager specifically for Invitrogen's LanthaScreen® Terbium-based TR-FRET Assay. It details the recommended optical components, including specific part numbers for excitation and emission filters and a dichroic mirror, to ensure compatibility and optimal performance.
| Excitation Filter Part Number | 14000101 |
|---|---|
| Excitation Filter Wavelength | 340/60 nm |
| Emission Filter 1 Part Number | 11440175 |
| Emission Filter 1 Wavelength | 525/20 nm |
| Dichroic Mirror Part Number | 10866181 |
| Recommended Plate Type | black opaque NBS plate |
No, it is unnecessary to establish a flat-field correction (FCC) when using LanthaScreen on the ViewLux.
A black opaque NBS plate is recommended, as white opaque plates may decrease the assay window and precision.