Siemens SITRANS F M MAGFLO Electromagnetic Flowmeters Handbook

This manual introduces the characteristics of the Siemens SITRANS F M MAGFLO series electromagnetic flowmeters, including models, sizes, connection methods, pressure, temperature, lining, electrodes, housing, explosion-proof version, and output.


Siemens SITRANS F M MAGFLO Electromagnetic Flowmeters Handbook - cover page
Brand
Siemens
Category
Other Electronics
Document type
User Manual
Language
English
Pages
80
File format
PDF
File size
7.4 MB
Published
27 October, 2023
Updated
27 October, 2023
MD5 checksum
1663D7603627F1C55DE5298792BB8AFF

About this manual

This handbook provides technical documentation for the complete range of SITRANS F M MAGFLO electromagnetic flowmeters, including sensor types MAG 1100, MAG 3100, and MAG 5100 W, and transmitter types MAG 5000 and MAG 6000. The document details the product introduction, mode of operation based on Faraday's law, and specific features such as the SENSORPROM memory unit and USM II communication modules.

What's inside

  • Product introduction
  • Mode of operation
  • Sensor types
  • Transmitter types
  • Sizing table
  • Installation conditions
  • Electrical connection
  • Commissioning
  • Troubleshooting
  • Ordering

Specifications

Flow directionUni/bidirectional
Meter uncertainty±0,5% o.r.
Input impedance>10 14 Ω
Turn down ratiominimum 3000:1
Power supply (MAG 5000/6000)12-24 V AC/DC

Frequently asked questions

How does the flowmeter resume measurement if the transmitter is replaced?

The new transmitter uploads all previous settings stored in the SENSORPROM unit and resumes measurement without any need for re-programming.

What types of liquids can these electromagnetic flowmeters measure?

These flowmeters are designed for measuring all electrically conductive liquids.

What is the minimum conductivity required for flow measurement?

The device allows flow measurements on fluids with conductivities as low as 1 mS/cm.

What is the minimum turn down ratio of the transmitter's dynamic range?

The transmitter has a turn down ratio of minimum 3000:1.