This is a manual about the operation of Shimadzu Multitype ICP Emission Spectrometer ICPE-9000 Hardware
The ICPE-9000 Hardware Instruction Manual from Shimadzu describes installation and operation procedures for the Multitype ICP Emission Spectrometer, along with precautionary information and details on accessories and options. It begins with before-use guidance, warranty and after-sales service information, replacement parts availability, and safety precautions for installation, maintenance, and repairs.
The manual includes a specifications section covering the plasma light source, RF power supply, spectrometer, and measuring unit. Stated specifications include an Echelle optical system, a wavelength range of 167 nm to 800 nm, a 27.12 MHz RF frequency, and a two-dimensional CCD detector.
| Optical system | Echelle mounting |
|---|---|
| Wavelength range | 167 nm to 800 nm |
| RF frequency | 27.12 MHz |
| RF output | 1.6 kW max. |
| Detector | Two-dimensional CCD semiconductor detector |
| Pixel size | 20 µm × 20 µm |
| Plasma torch | Mini-torch made of quartz |
| Plasma gas flow rate range | 0 to 20 L/min |
The wavelength range is 167 nm to 800 nm.
Installation is performed by Shimadzu-designated service engineers.
The RF output is 1.6 kW max.
It uses a two-dimensional CCD semiconductor detector.