This document provides standard operating procedures, maintenance guidelines, troubleshooting steps, and specifications for Shimadzu AP series electronic…
The manual details the working principles of the Shimadzu AP series electronic analytical balance, which utilizes a UniBloc AP electromagnetic force restoration sensor. It explains that temperature changes can affect measurement stability due to thermal expansion in components like coils and circuit boards, necessitating sufficient warm-up time before use.
The guide covers operational procedures including leveling the instrument, performing automatic or manual calibration using internal or external weights, and various weighing functions such as zeroing, taring (peeling), and adding materials. It also provides specific maintenance tips for cleaning stainless steel parts with anhydrous ethanol and avoiding organic solvents on the display. This manual is written in Chinese.
| Model AP135W Range | 135g |
|---|---|
| Model AP125WD Precision | 0.1mg/0.01mg |
| Model AP324X Range | 320g |
| Warm-up Time (1/100,000) | Four hours or more |
It requires warm-up for four hours or more after being powered on.
Anhydrous ethanol should be used to clean stainless steel parts like the weighing pan regularly.
It indicates that the object being weighed is overloaded.