Standard operating procedure for water total organic carbon analysis using the Shimadzu TOC-V CSN and ASI-V autosampler.
This document is a Standard Operating Procedure (SOP) from the California Department of Pesticide Regulation detailing the use of the Shimadzu TOC-V CSN analyzer with the ASI-V autosampler for measuring Total Organic Carbon in water samples. It outlines necessary materials, including specific reagents like 2M HCl and low TOC water.
The needle of the left gauge should be between 44 -87 psi; 60 psi is the optimum operating pressure.
The bottom of the ball should be level to the 150 mL /min increment in the scale.
Fill the glass vials with approximately 35 mL of water with the meniscus just below the shoulders of the vial.