This document introduces the basic safety information for using Beckman Coulter 50.2 Ti rotor.
This manual provides safety guidelines, technical specifications, and operational procedures for the Beckman Coulter 50.2 Ti rotor. It details handling, loading, and maintenance requirements, including safety precautions for biological materials and speed limitations. The rotor is designed for use with Beckman Coulter ultracentrifuges and is rated for 50,000 rpm with a maximum capacity of 468 mL. It is made of titanium and includes components like O-rings and an overspeed detection disk for safety and performance.
| Maximum speed | 50 000 rpm |
|---|---|
| Density rating at maximum speed | 1.2 g/mL |
| Relative Centrifugal Field at rmax | 302 000 × g |
| k factor at maximum speed | 69 |
| Number of tube cavities | 12 |
| Nominal tube capacity (largest tube) | 39 mL |
| Nominal rotor capacity | 468 mL |
| Rotor material | titanium |
The maximum speed is 50,000 rpm.
The rotor is made of titanium and is finished with black polyurethane paint.
The rotor has a nominal capacity of 468 mL.