This product is a steam sterilizer, which is used to sterilize medical devices, dental devices, and surgical instruments.
This manual provides step-by-step instructions for operating SUN series autoclaves, including setup, sterilization programs, and troubleshooting. It details technical parameters, installation requirements, and maintenance procedures. The document also includes a section on emergency conditions with error codes and their resolutions, ensuring safe operation.
The autoclave is designed for frequent sterilization in dental, ophthalmology, and lab settings. It features a microprocessor-controlled system with an LCD display and keypad for intuitive operation. The manual covers filling and draining water, adjusting settings, and cleaning procedures to maintain performance and safety.
| Chamber Size (18L) | 249mmX355mm |
|---|---|
| Rated Voltage | 220V±22V, 50Hz±1Hz or 110V±11V, 60Hz±1Hz |
| Rated Power | 1800VA 2000VA |
| Sterilizing Pressure/Temp | 100-120KPa/121ºC or 200-220KPa/134ºC |
| Water Tank Capacity | 2L |
| Operating Temperature Range | 0-40ºC |
The sterilization programs include: 134℃ UNWRAPPED, 121℃ UNWRAPPED, 134℃ WRAPPED, 121℃ WRAPPED, 134℃ COTTON, 121℃ COTTON, and 121℃ PLASTIC.
Fill the tank with distilled water using a measuring cup. Fill approximately 1500ml for the first time, then refill as needed. Drain used water after sterilization and clean the brass filter inside the chamber.
Check the error code table for the specific malfunction and follow the resolving measures. If unable to resolve, contact the dealers for assistance.
Clean the interior surface of the autoclave chamber monthly with disinfectant or alcohol.