This document provides instructions for calibrating and maintaining the 7300Plus Real-Time PCR instrument.
The Applied Biosystems 7300Plus Real-Time PCR Instrument is a 96-well, four-color platform designed for quantitative and qualitative detection of nucleic acid sequences using fluorescence-based polymerase chain reaction (PCR) reagents. This user guide details the procedures for calibrating the instrument, including Regions of Interest (ROI), background, optical, and dye calibrations, as well as verifying instrument performance using an RNase P experiment.
The document also covers user-performed maintenance tasks such as monitoring the instrument status, decontaminating the sample block, replacing the halogen lamp, and updating the Windows Operating System and Real-Time PCR Software. It includes safety information, storage and installation instructions, and guidance on creating custom dye and background plates.
| Well configuration | 96-well |
|---|---|
| Detection method | fluorescence-based polymerase chain reaction |
| Software version | v2.4 or later |
The instrument is for research use only and is not for use in diagnostic procedures.
The instrument should be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
The instrument uses FAM, SYBR, JOE, VIC, TAMRA, NED, Cy3, ROX, Texas Red dyes.