The Applied Biosystems ™ 3730xl DNA Analyzer with 3730xl Data Collection Software 5 is a fluorescence-based DNA analysis instrument. It uses capillary electrophoresis technology with 48 or 96 capillaries.
This document serves as a Quick Reference for the Applied Biosystems 3730xl DNA Analyzer equipped with 3730xl Data Collection Software 5 and Windows 10. It provides essential instructions for powering on the system, filling water and buffer reservoirs, and replacing the anode buffer jar. The guide emphasizes safety precautions, such as wearing protective eyewear and cleaning reservoirs weekly.
| Technology | fluorescence-based DNA analysis |
|---|---|
| Capillary count | 48 or 96 capillaries |
| Software version | 3730xl Data Collection Software 5 |
Replace the anode buffer before each group of scheduled runs, at least every 48 hours, or after replenishing polymer.
Discard old buffer and add fresh buffer; using old buffer can lead to loss of resolution and data quality.
Clean the buffer, water, waste, and buffer reservoirs weekly in warm water followed by a rinse with deionized water.