Applied Biosystems ViiA 7 Real-Time PCR System Quick Reference Guide

The ViiA™ 7 Real-Time PCR System uses fluorescent-based polymerase chain reaction (PCR) reagents to provide quantitative research detection of target nucleic acid sequences (targets) and qualitative research detection of targets using post-PCR (endpoint) analysis.


Applied Biosystems ViiA 7 Real-Time PCR System Quick Reference Guide - cover page
Brand
Applied Biosystems
Category
Other
Document type
Quick Start Guide
Language
English
Pages
20
File format
PDF
File size
744 KB
Published
21 October, 2023
Updated
21 October, 2023
MD5 checksum
5F10ACCAADABD269A3DBB6A6D9FC4243

About this manual

This document serves as a quick reference for the Applied Biosystems ViiA 7 Real-Time PCR System, detailing the use of QuantStudio Real-Time PCR Software v1.6.1 for calibration. It outlines specific calibration requirements for different plate blocks, such as the 96-well and 384-well blocks, and notes that normalization calibration is only required for the TaqMan Array Card block after a software upgrade.

The guide provides a recommended schedule for user-performed maintenance tasks, ranging from weekly checks of computer disk space and instrument self-tests to semi-annual calibrations like regions of interest (ROI), background, uniformity, and dye calibrations. It also lists specific calibration and verification consumables, including spectral calibration kits and verification plates for various well formats.

What's inside

  • Calibrate and maintain the ViiA 7
  • Recommended calibration and maintenance schedule
  • Calibration and verification consumables
  • Prepare the calibration plate or array card
  • Safety guidelines
  • Software version requirements
  • Plate block calibration rules
  • Consumable storage and shelf life

Specifications

Software Versionv1.6.1
Publication NumberMAN0018831 Rev. C.0
Plate Block Types96-well, 384-well, TaqMan Array Card

Frequently asked questions

Does the ViiA 7 require normalization calibration for 96-well plates?

No, QuantStudio Real-Time PCR Software v1.6.1 does not require a normalization calibration for the 96-well plate blocks.

When is normalization calibration required?

Normalization calibration is only required for the TaqMan Array Card block with this version of the software.

What should be done if the software is upgraded?

The instrument must be calibrated before starting a run with a 96-well plate block or a 384-well plate block.