The ViiA™ 7 Real-Time PCR System uses fluorescent-based polymerase chain reaction (PCR) reagents to provide quantitative research detection of target nucleic acid sequences (targets) and qualitative research detection of targets using post-PCR (endpoint) analysis.
This document serves as a quick reference for the Applied Biosystems ViiA 7 Real-Time PCR System, detailing the use of QuantStudio Real-Time PCR Software v1.6.1 for calibration. It outlines specific calibration requirements for different plate blocks, such as the 96-well and 384-well blocks, and notes that normalization calibration is only required for the TaqMan Array Card block after a software upgrade.
The guide provides a recommended schedule for user-performed maintenance tasks, ranging from weekly checks of computer disk space and instrument self-tests to semi-annual calibrations like regions of interest (ROI), background, uniformity, and dye calibrations. It also lists specific calibration and verification consumables, including spectral calibration kits and verification plates for various well formats.
| Software Version | v1.6.1 |
|---|---|
| Publication Number | MAN0018831 Rev. C.0 |
| Plate Block Types | 96-well, 384-well, TaqMan Array Card |
No, QuantStudio Real-Time PCR Software v1.6.1 does not require a normalization calibration for the 96-well plate blocks.
Normalization calibration is only required for the TaqMan Array Card block with this version of the software.
The instrument must be calibrated before starting a run with a 96-well plate block or a 384-well plate block.