This document introduces the usage and precautions of the Z6Ms True Volume TEE Transducer, including the storage, transportation, accident prevention, conductivity and leakage testing, cleaning methods, etc.
This document serves as a care guide for the Siemens Z6Ms True Volume TEE Transducer, covering essential procedures for storage, transport, and accident prevention. It outlines the correct handling of the device before use, including inspection for physical damage and the mandatory performance of conductivity and leakage tests using the Dale 800B Tester.
| Test Depth Limit | 100 cm |
|---|---|
| Handle Submersion | Not allowed |
Test again, and if it fails, do not use the transducer and call the Service Representative.
The electrode rods must be placed into the solution at a minimum depth of 1 inch (25mm).
No, the handle cannot be submerged into the liquid.