M1 Mini User Manual introduces the precautions for using the device, the model, function, product features and other information.
This user manual provides step-by-step instructions for setting up and operating the Philips M1 Mini Auto CPAP system. It includes guidance on downloading the LightTrip app, installing filters, connecting tubing and masks, adjusting airflow, and navigating patient settings. The manual also covers cleaning, travel use, device transfer, troubleshooting, and technical specifications.
Safety warnings and contraindications are clearly outlined, including restrictions for use in MRI environments and with electromagnetic equipment. The device is intended for adult use only and must be prescribed by a healthcare professional. Maintenance and cleaning procedures are provided for mask, tubing, and enclosure components.
| Device Size Dimensions | 159 mm × 66 mm × 72 mm |
|---|---|
| Weight | < 400 g |
| Operating Temperature | 5°C to 35°C (41˚F to 95˚F) |
| Humidity | ≤ 93% Non-condensing |
| Atmospheric Pressure | 760 ~ 1060 hPa |
| AC Power Consumption | 100 - 240 V AC, 50 / 60 Hz, 1.0 A max |
| Main device input | 19 V, 1.26 A |
| Pressure Range | 4 to 20 hPa (in 0.5 hPa increments), ≤ 30 hPa under single fault conditions. |
| Pressure Display Accuracy | 0 to 20 hPa,Margin of Error:± (0.5 hPa+4%) |
| Static Pressure Stability | ± 0.5 hPa |
| Ramp Time | 0 to 60 minutes |
| Sound Pressure Level | < 30 dB (at 10 hPa) |
| Sound Power Level | < 38 dB (at 10 hPa) |
| Air Tubing Inner Diameter | 15 mm and 19 mm |
| Patient Connection Port | 22 mm conical air outlet |
| Air Filter Filtration Efficiency | > 20% for 10 micron |
| Bluetooth Product Name | Auto CPAP System |
| Bluetooth Model | M1 Mini |
The M1 Mini Auto CPAP system is designed for the treatment of adults with Obstructive Sleep Apnea (OSA) only, either in the hospital or at home.
Absolute contraindications include pneumothorax, cerebrospinal fluid leak, shock, active epistaxis, upper gastrointestinal bleeding, coma, or giant vocal fold polyp. Relative contraindications include severe coronary heart disease, acute otitis media, excessive secretions, weak cough, tracheal intubation, severe nasal congestion, lung bullae, or allergies to masks.
The device can operate and be stored between 5°C to 35°C (41°F to 95°F) for operation and -25°C to 70°C (-13°F to 158°F) for storage.
The pressure range is 4 to 20 hPa (in 0.5 hPa increments), with a maximum of 30 hPa under single fault conditions.