The document is the instruction manual for Perma Pure MDH-Series Gas Dryer
The Perma Pure MDH-Series gas dryer utilizes proprietary Nafion technology to selectively remove water from gas samples without affecting other components. The heated design allows it to handle gas samples with dew points higher than ambient temperature, preventing condensation without the need for separate heating elements or insulation. It is designed for compact integration into analyzer cabinets or standalone use.
| Max Gas Flow | 1.5 lpm |
|---|---|
| Water Removal Capacity | 30% by volume |
| Thermostat Set Point | 80° C |
| Power Requirement (AC) | 110 VAC, 0.3 Amps, 30 Watts |
| Power Requirement (DC) | 12 VDC, 2.5 Amps |
| Max Sample Pressure | 80 psig |
| Max Purge Pressure | 10 psig |
It uses proprietary Nafion technology to selectively remove only water from sample gases, leaving the rest of the gas sample unaffected.
Purge air must be at -40°C dew point, flow at 15 times the sample flow rate, and flow countercurrent to the sample.
Decrease or stop the sample flow and increase the purge flow until the unit dries out.