Perma Pure MDH-Series Gas Dryer User Manual

The document is the instruction manual for Perma Pure MDH-Series Gas Dryer


Perma Pure MDH-Series Gas Dryer User Manual - cover page
Brand
Perma Pure
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
966 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
DB9CE2A9D571BFB70FCA98BA5B186488

About this manual

The Perma Pure MDH-Series gas dryer utilizes proprietary Nafion technology to selectively remove water from gas samples without affecting other components. The heated design allows it to handle gas samples with dew points higher than ambient temperature, preventing condensation without the need for separate heating elements or insulation. It is designed for compact integration into analyzer cabinets or standalone use.

What's inside

  • PRINCIPLE OF OPERA TION
  • INSTALLATION GUIDELINES
  • TROUBLESHOOTING

Specifications

Max Gas Flow1.5 lpm
Water Removal Capacity30% by volume
Thermostat Set Point80° C
Power Requirement (AC)110 VAC, 0.3 Amps, 30 Watts
Power Requirement (DC)12 VDC, 2.5 Amps
Max Sample Pressure80 psig
Max Purge Pressure10 psig

Frequently asked questions

What is the principle of operation for the MDH-Series gas dryer?

It uses proprietary Nafion technology to selectively remove only water from sample gases, leaving the rest of the gas sample unaffected.

What are the installation guidelines for purge air?

Purge air must be at -40°C dew point, flow at 15 times the sample flow rate, and flow countercurrent to the sample.

What should be done if condensation forms in the unit?

Decrease or stop the sample flow and increase the purge flow until the unit dries out.