This manual covers unpacking, installation, electrical connections, preheating, system fine tuning, maintenance and replacement parts for the Permapure Mini-GASS™ Sample Conditioning System.
This user manual provides instructions for installing, operating, and maintaining the Mini-GASS™ Sample Conditioning System by Perma Pure. It includes safety warnings, specifications, electrical and plumbing connections, startup procedures, and maintenance guidelines. The system uses membrane technology to dry gas samples and is designed for industrial applications requiring precise moisture control.
The manual also covers optional components like heated lines, purge air systems, steam heating, and filter drains, along with diagrams and troubleshooting for specific models such as MG-1228, MG-2812P, and MG-1220.
| Sample flow rate | 0 to 10 lpm |
|---|---|
| Sample inlet temperature | 250°F (121°C) max. with stainless steel filter |
| Sample inlet water vapor | 30% by volume maximum |
| Sample inlet pressure | Without purge eductor option: maximum 20 psig; minimum 5” H2O vacuum |
| Purge air requirements | Maximum 1 cfm of oil-free instrument air or nitrogen with less than -40°C dew point |
| Outlet sample dew point (PD-200T-24SS) | 26° F (-4° C) at 10 lpm |
| Electrical requirements | 115V ±10% or 230V ±10%, 50/60Hz, 8.0A/4.0A (up to 900 Watts depending on options) |
| Enclosure | NEMA 4X, Fiberglass with polycarbonate cover |
| Dimensions (MG-1228) | 12” W x 28” H x 7” Deep |
| Operating environment | -20°C to 40°C ambient temperature; 0-95% RH |
Maximum 250°F (121°C) with stainless steel filter.
26° F (-4° C) at 10 lpm.
115V ±10% or 230V ±10%, 50/60Hz, 8.0A/4.0A (up to 900 Watts depending on options).
NEMA 4X, Fiberglass with polycarbonate cover.