Perma Pure Mini-GASS™ Sample Conditioning System User’s Manual

This manual explains how to use the mini-gass sample conditioning system, including system installation, electrical connections, start-up procedures, preheating the system, system fine tuning, maintenance and replacement parts.


Perma Pure Mini-GASS™ Sample Conditioning System User’s Manual - cover page
Brand
Perma Pure
Category
Analyzers
Document type
User Manual
Language
English
Pages
30
File format
PDF
File size
1.4 MB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
7041CAC4D77783CB71A121D102C62CFC

About this manual

This manual provides instructions for installing, operating, and maintaining the Mini-GASS™ Sample Conditioning System by Perma Pure. It includes safety warnings, specifications, electrical and plumbing connections, and maintenance procedures. The system uses membrane technology to dry gas samples and requires proper ventilation for purge exhaust to ensure safe operation.

The system comes with options like heated enclosures, filters, and dryers, and must be installed by trained personnel following local safety codes and electrical standards. It is designed for use in dry environments and requires oil-free instrument air or nitrogen with a low dew point for optimal performance.

What's inside

  • Unpacking
  • Important safety warnings
  • General description
  • Specifications and Features
  • Installation
  • Electrical connections
  • Start-up procedure
  • Preheating the system
  • System fine tuning
  • Maintenance

Specifications

Sample flow rate0 to 10 lpm
Sample inlet temperature250°F (121°C) max. with stainless steel filter
Sample inlet water vapor30% by volume maximum non condensing
Purge air requirementsMaximum 1 cfm of oil-free instrument air or nitrogen with less than -40°C dew point
Electrical requirements115V ±10% or 230V ±10%, 50/60Hz, 5.0A/2.5A
EnclosureNEMA 4X, Fiberglass with polycarbonate cover
Operating environment-20°C to 40°C ambient temperature; 0-95% RH

Frequently asked questions

What is the maximum sample inlet temperature?

250°F (121°C) maximum with a stainless steel filter.

What type of purge air is required?

Oil-free instrument air or nitrogen with a dew point no higher than -40°C.

What is the outlet sample dew point with the PD-200T-24SS dryer at 10 lpm?

26°F (-4°C).

How should the purge exhaust be vented?

It must be permanently routed to a vent system or directly outdoors; do not vent into a structure or confined space.