This manual explains how to use the mini-gass sample conditioning system, including system installation, electrical connections, start-up procedures, preheating the system, system fine tuning, maintenance and replacement parts.
This manual provides instructions for installing, operating, and maintaining the Mini-GASS™ Sample Conditioning System by Perma Pure. It includes safety warnings, specifications, electrical and plumbing connections, and maintenance procedures. The system uses membrane technology to dry gas samples and requires proper ventilation for purge exhaust to ensure safe operation.
The system comes with options like heated enclosures, filters, and dryers, and must be installed by trained personnel following local safety codes and electrical standards. It is designed for use in dry environments and requires oil-free instrument air or nitrogen with a low dew point for optimal performance.
| Sample flow rate | 0 to 10 lpm |
|---|---|
| Sample inlet temperature | 250°F (121°C) max. with stainless steel filter |
| Sample inlet water vapor | 30% by volume maximum non condensing |
| Purge air requirements | Maximum 1 cfm of oil-free instrument air or nitrogen with less than -40°C dew point |
| Electrical requirements | 115V ±10% or 230V ±10%, 50/60Hz, 5.0A/2.5A |
| Enclosure | NEMA 4X, Fiberglass with polycarbonate cover |
| Operating environment | -20°C to 40°C ambient temperature; 0-95% RH |
250°F (121°C) maximum with a stainless steel filter.
Oil-free instrument air or nitrogen with a dew point no higher than -40°C.
26°F (-4°C).
It must be permanently routed to a vent system or directly outdoors; do not vent into a structure or confined space.