eCOOL™ 5000 Series Digital Thermo-Electric Cooler Instruction Manual
This manual provides detailed instructions for installing, operating, and maintaining the Perma Pure eCOOL™ 5000 Series digital thermo-electric coolers. It covers specifications for various models (e5300 to e5900XP), installation guidelines, start-up procedures, LED and LCD indicators, board configurations, default setpoints, thermocouple options, maintenance, troubleshooting, and spare parts. The coolers use Peltier elements to cool sample gas to a desired dew point, with features like remote monitoring, alarms for temperature and moisture issues, and compatibility with multiple power supplies.
The manual includes technical specifications such as heat exchanger dimensions, power requirements, and operating parameters like maximum inlet temperature and pressure. It also explains the principle of operation, including how the Peltier elements work and how the cooling system manages heat transfer. Installation details address tubing connections, fitting types, and precautions to prevent damage to components. Maintenance and troubleshooting sections guide users through common issues and solutions.
| Model | e5300, e5400, e5500, e5500XP, e5800, e5800XP, e5900, e5900XP |
|---|---|
| Heat Exchanger Material | Stainless steel (standard); Durinert® treated stainless steel, Kynar®, Glass (optional) |
| Power Supply | 110/120 VAC, 220/240 VAC, 12 VDC, 15 VDC, 24 VDC |
| Maximum Inlet Sample Temperature | 400°F (205°C) for stainless steel, Durinert®, Glass; 280°F (138°C) for Kynar® |
| Maximum Inlet Pressure | 45 psig (3 bar / 2250 mmHg) |
| Ambient Temperature Range | 33-104°F (0.6-40°C) |
| Outlet Sample Gas Dew Point | 41°F (5°C) |
The coolers operate with 110/120 VAC, 220/240 VAC, 12 VDC, 15 VDC, or 24 VDC.
The maximum inlet sample temperature is 400°F (205°C).
The ambient temperature range is 33-104°F (0.6-40°C).
The outlet sample gas dew point is 41°F (5°C).