Instruction manual for nXDS6i/10i/15i/20i/6iC/10iC/15iC/20iC/6iR/10iR/15iR/20iR series products
This manual provides operational instructions for Edwards nXDS series serial communication interfaces. It details how to connect and operate the devices using RS232 or RS485 serial protocols. The document covers control modes, serial enable requirements, message structure, and command sets for pump operation. It also includes technical data for logic interface pins and voltage specifications.
The manual is intended for users who need to configure and operate these interfaces for serial control and monitoring. It includes diagrams for connections and tables for pin assignments and technical parameters. Safety and operating information are referenced in the main product instruction manual.
| Power Supply | 100-127/200-240 V, 1 phase, 50/60 Hz |
|---|---|
| Interface Type | RS232 or RS485 |
| Baud Rate | 9600 Baud |
| Data Format | 8 bits, 1 stop, no parity |
| Maximum Message Size | 80 characters |
| Logic Interface Connector | 15-way D-type (male) |
| Analogue Speed Input | 0-10 V d.c. |
| FAIL Status Output | Open collector transistor plus pull up resistor |
| Analogue Reference Output | +10 V d.c. unipolar output |
The required power supply is 100-127/200-240 V, 1 phase, 50/60 Hz.
The pump supports RS232 and RS485 serial protocols.
The baud rate is 9600 Baud, with 8 bits, 1 stop, no parity.
Connect the serial enable input (pin 5) to pin 2 of the logic interface mating half.