This instruction manual covers installation, operation, and maintenance for Edwards LCPV16EKA and LCPV25EKA solenoid isolation valves.
This manual provides installation, operation, and maintenance instructions for the Edwards LCPV16EKA and LCPV25EKA solenoid isolation valves. These solenoid-operated valves are designed for general vacuum use to open and close a vacuum line, utilizing a fluoroelastomer pad on the plunger base to seal against the valve body.
| Mass (LCPV25EKA) | 840 g |
|---|---|
| Ambient operating temperature | 5 to 45 ºC |
| Vent gas temperature range | -10 to 130 ºC |
| Maximum altitude | 4000 m |
| Leak rate | < 1 x 10 -6 mbar.ls-1 |
| Time to open | 60 msec |
| Time to close | < 20 msec |
| Maximum cycle frequency | 1800/hour |
| Reliability (MTTF) | 500,000 cycles |
| Enclosure rating | IP67 |
| Body material | Aluminium |
| Valve seal material | Fluoroelastomer |
| Actuator material | Stainless steel |
| Coil insulation class | Class H |
No, the LCPVEKA valves do not require routine maintenance during their normal operational life and cannot be serviced by the user.
The valve closes automatically when the electrical supply is switched off.
The maximum cycle frequency is 1800 cycles per hour.