Edwards STP-603/1003 Series Turbomolecular Pump Instruction Manual

This is an instruction manual for STP pump, which mainly introduces the characteristics and usage methods of STP pump.


Edwards STP-603/1003 Series Turbomolecular Pump Instruction Manual - cover page
Brand
Edwards
Category
Industrial Robot
Document type
Service Manual
Language
English
Pages
23
File format
PDF
File size
691 KB
Published
17 October, 2023
Updated
17 October, 2023
MD5 checksum
52CBFDB43F6224ADF347AB09CF60F432

About this manual

This manual provides specific installation and securing instructions for the Edwards STP-603/1003 Series turbomolecular pumps. It includes technical specifications, flange sizes, pumping speeds, and cooling methods. The document also details how to secure the inlet port using bolts, claw clamps, or dampers. It is part of a three-volume set, with this volume focusing on pump-specific information.

The pump is designed for ultra-high vacuum applications and requires a compatible control unit (SCU-750/SCU-800). It operates with natural air cooling and has a maximum working pressure of 41.3×10⁵ Pa (1104 Torr). The manual includes illustrations and tables for tightening torques and flange clamp specifications.

What's inside

  • Technical Data
  • How to Secure the STP Pump
  • Flange Sizes
  • Installation
  • Temperature Management System (TMS)
  • Accessories
  • Maintenance
  • Storage and Disposal

Specifications

Flange SizeICF203/VG150 for 603 series; ICF253/VG200/ISO200/ISO250/ISO250F for 1003 series
Pumping Speed650 L/s for 603 series; 1000 L/s for 1003 series
Ultimate Pressure (Non-corrosion)10⁻⁷ Torr (order: ICF flange)
Maximum Working Pressure41.3×10⁵ Pa (1104 Torr)
Rated Speed35,000 rpm
Noise Level<50 dB (at 35,000 rpm)
Mass31 kg
Ambient Temperature Range0 to 40°C

Frequently asked questions

What control units are compatible with the STP-603/1003 series?

The compatible control units are SCU-750 and SCU-800.

What is the ultimate pressure for the non-corrosion type?

The ultimate pressure is 10⁻⁷ Torr (order: ICF flange).

What is the maximum working pressure?

The maximum working pressure is 41.3×10⁵ Pa (1104 Torr).

How is the pump cooled?

The pump uses natural air cooling, with options for water or air cooling during baking and gas pumping.

Download PDF (691 KB)