This document is an instruction manual for the MT-57E-101-A Voltage SCU-XL800 Control Unit. The control unit is suitable for 200-240V AC and 100-120V AC. The document consists of three volumed instruction manuals for the STP pump, including a generic instruction manual, specific information manual, and control unit manual. Each manual covers installation, operation, maintenance, storage, and other topics.
This document serves as the third volume of a three-part instruction set for the Edwards SCU-XL800 Control Unit, specifically designed for use with STP series turbomolecular pumps. It provides detailed guidance on the unit's introduction, technical specifications, and installation procedures, including rack mounting and cable connections for power, TMS, and STP units.
| Applicable pump | STP-603/1003 series, STP-A803/A1303 series |
|---|---|
| Input voltage | 100 to 120 Va.c. ± 10%, 200 to 240 Va.c. ± 10% |
| Input power (without TMS) | 850 VA maximum |
| Input power (with TMS) | 1200 VA maximum |
| Input frequency | 50/60 ± 2 Hz |
| Leakage current | 3.5 mA maximum |
| Main breaker rated current | 15 A |
| Interrupting Capacity | 1000 A (240 Va.c., 50/60 Hz) |
| Output voltage | 92 Va.c. maximum |
| Output frequency | 800 Hz maximum |
| Allowable ambient temperature | 0 to 40 °C (32 to 104 °F) |
| Storage temperature | -25 to 55 °C (-13 to 131 °F) |
| Mass | 11 kg (24.2 lb) |
| TMS Battery | Lithium battery, 3 V, 130 mAhr |
| Serial communication | RS232/RS485 |
The warranty period is not explicitly stated in the provided text, but the document includes a section on limited warranty.
The unit includes detection for electromagnetic bearing failure, STP pump overheat, motor driver overload, power failure, overspeed, control unit overheat, TMS unit failure, and air-cooling fan failure.
It should be disposed of as industrial waste in accordance with all local and national safety and environmental standards.