MT-57E-101-A Voltage SCU-XL800 Control Unit Instruction Manual

This document is an instruction manual for the MT-57E-101-A Voltage SCU-XL800 Control Unit. The control unit is suitable for 200-240V AC and 100-120V AC. The document consists of three volumed instruction manuals for the STP pump, including a generic instruction manual, specific information manual, and control unit manual. Each manual covers installation, operation, maintenance, storage, and other topics.


MT-57E-101-A Voltage SCU-XL800 Control Unit Instruction Manual - cover page
Brand
Edwards
Category
Industrial Supply
Document type
User Manual
Language
English
Pages
131
File format
PDF
File size
2 MB
Published
15 October, 2022
Updated
28 September, 2023
MD5 checksum
3A56CB33550975FCF3C5C8214003BE1F

About this manual

This document serves as the third volume of a three-part instruction set for the Edwards SCU-XL800 Control Unit, specifically designed for use with STP series turbomolecular pumps. It provides detailed guidance on the unit's introduction, technical specifications, and installation procedures, including rack mounting and cable connections for power, TMS, and STP units.

What's inside

  • Introduction
  • Technical Data
  • Installation
  • Operation
  • Serial Communication Protocol
  • STP-Link
  • Maintenance
  • Storage and Disposal
  • Service and Spares

Specifications

Applicable pumpSTP-603/1003 series, STP-A803/A1303 series
Input voltage100 to 120 Va.c. ± 10%, 200 to 240 Va.c. ± 10%
Input power (without TMS)850 VA maximum
Input power (with TMS)1200 VA maximum
Input frequency50/60 ± 2 Hz
Leakage current3.5 mA maximum
Main breaker rated current15 A
Interrupting Capacity1000 A (240 Va.c., 50/60 Hz)
Output voltage92 Va.c. maximum
Output frequency800 Hz maximum
Allowable ambient temperature0 to 40 °C (32 to 104 °F)
Storage temperature-25 to 55 °C (-13 to 131 °F)
Mass11 kg (24.2 lb)
TMS BatteryLithium battery, 3 V, 130 mAhr
Serial communicationRS232/RS485

Frequently asked questions

What is the warranty period for the SCU-XL800 Control Unit?

The warranty period is not explicitly stated in the provided text, but the document includes a section on limited warranty.

What are the safety functions of the SCU-XL800?

The unit includes detection for electromagnetic bearing failure, STP pump overheat, motor driver overload, power failure, overspeed, control unit overheat, TMS unit failure, and air-cooling fan failure.

How should the SCU-XL800 be disposed of?

It should be disposed of as industrial waste in accordance with all local and national safety and environmental standards.

Download PDF (2 MB)