Ohaus MB23 and MB25 Moisture Analyzers Service Manual

This document is a user manual for MB23 and MB25 Moisture Analyzers, which contains instructions on how to operate and maintain the device, as well as troubleshooting methods


Ohaus MB23 and MB25 Moisture Analyzers Service Manual - cover page
Brand
OHAUS
Category
Other
Document type
Service Manual
Language
English
Pages
53
File format
PDF
File size
1.7 MB
Published
19 October, 2023
Updated
19 October, 2023
Models
MB23, MB25
MD5 checksum
A99C01E9906744A90F1A7B48B2CCD7C0

About this manual

This document is a service manual for the Ohaus MB23 and MB25 Moisture Analyzers. It provides instructions for routine maintenance, disassembly, repair, and replacement of internal components such as the power PCB, main PC board, load cell, heating unit, and fan. The manual assumes familiarity with the user instruction manual and covers safety requirements for the service area.

What's inside

  • Getting Started
  • Troubleshooting
  • Maintenance Procedures
  • Final Testing
  • Drawings and Parts Lists
  • Standard Calibration
  • Service Calibration
  • Software Service Tool

Specifications

Repeatability (Std Dev)0.2% (3g sample) / 0.05% (10g sample) (MB25)
Readability0.005g; 0.05% (MB25)
Operating Temperature Range41°F to 104°F / 5°C to 40°C
Power Requirement100 to 240 VAC, 50/60 Hz

Frequently asked questions

What is the stabilization time for the analyzer?

The stabilization time is less than or equal to 3 seconds.

What warm-up time is required before performing performance tests?

You must allow the Moisture Analyzer to warm up for at least one hour before testing.

What class of tolerance mass is required for performance tests?

An ASTM Class 1 or OIML Class E2 Tolerance mass should be used.

What is the maximum stabilization time for the MB23 and MB25?

The stabilization time is less than or equal to 3 seconds.