This is a document about the calibration of Ohaus Valor™ 1000 weighing scale, which introduces the calibration method of the scale, including preparations before calibration and calibration steps
This document serves as a calibration guide for the Ohaus Valor 1000 weighing scale. It outlines the standard span calibration process, noting that specific steps vary depending on whether the device is running software version 1.xx or version 2.xx and later. Users must ensure the scale is level and stable, and allow it to warm up for 20 minutes before beginning the procedure.
| Calibration Weight | 3.000kg or 3000.0g |
|---|---|
| Warm-up Time | 20 minutes |
The standard calibration process is Span calibration.
To abort calibration, power off the scale.
The span weight is 3000 g (3 kg) for all models.