This document provides supplemental information for the Nellcor OXIMAX N-65 Pulse Oximeter
This document provides supplemental instructions for the Nellcor OXIMAX N-65 Pulse Oximeter, including safety details, alarm settings, and signal processing behavior. It explains how the device handles low perfusion, dynamic averaging, and compliance with ISO 9919:2005 standards. It also addresses testing limitations and compatibility requirements for accurate SpO2 measurements.
| Peak Wavelengths | Red: 660 nm, Infrared: 900 nm |
|---|---|
| Maximum Output Power | Less than 15 mW |
| Default SpO2 Alarm Limit (Neonatal Mode) | 80% |
| Data Update Period (Normal Conditions) | 6 to 7 seconds (3 seconds in Fast Mode) |
The total optical output power is less than 15 mW.
The PULSE SEARCH indicator illuminates solid, and SpO2 and pulse rate update every second.
SpO2 accuracy can only be validated in vivo by comparing pulse oximeter readings with values traceable to SaO2 measurements from a laboratory CO-oximeter.