Nellcor N-65 Pulse Oximeter Addendum to Monitor Instructions

This document provides supplemental information for the Nellcor OXIMAX N-65 Pulse Oximeter


Nellcor N-65 Pulse Oximeter Addendum to Monitor Instructions - cover page
Brand
Nellcor
Category
Health Care
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
48 KB
Published
19 October, 2023
Updated
19 October, 2023
Model
N-65
MD5 checksum
1FA5A53EA135229512DCCECBC04838A5

About this manual

This document provides supplemental instructions for the Nellcor OXIMAX N-65 Pulse Oximeter, including safety details, alarm settings, and signal processing behavior. It explains how the device handles low perfusion, dynamic averaging, and compliance with ISO 9919:2005 standards. It also addresses testing limitations and compatibility requirements for accurate SpO2 measurements.

What's inside

  • Safety
  • Alarms
  • Range of Peak Wavelengths and Maximum Output Power
  • Data Update Period, Effect Of Data Averaging, and Other Signal Processing
  • Test Considerations and Oximeter Accuracy
  • Compliance

Specifications

Peak WavelengthsRed: 660 nm, Infrared: 900 nm
Maximum Output PowerLess than 15 mW
Default SpO2 Alarm Limit (Neonatal Mode)80%
Data Update Period (Normal Conditions)6 to 7 seconds (3 seconds in Fast Mode)

Frequently asked questions

What is the maximum output power of the N-65 sensor LEDs?

The total optical output power is less than 15 mW.

What happens if the dynamic averaging time for SpO2 exceeds 20 seconds?

The PULSE SEARCH indicator illuminates solid, and SpO2 and pulse rate update every second.

What is required to validate SpO2 accuracy?

SpO2 accuracy can only be validated in vivo by comparing pulse oximeter readings with values traceable to SaO2 measurements from a laboratory CO-oximeter.

Download PDF (48 KB)